Download Documentation - Tech
Transcript
4.5.5 Room temperature measurement 4.5.5.1 Normal force Analog Input 0 → high temperature normal force Due to the different geometries with the room temperature force module installed, the strain gauges measuring bridge has another characteristic curve for the normal force. The corresponding load of the output voltages (carrier frequency amplifier) are measured with the already calibrated room temperature force module and calculated with its characteristic linear function. Carrier frequency amplifiers channel 3 settings for calibration: • Set the basic parameter of the measuring range to 0.5 mV/V characteristic curve Fn room temperature 200.00 180.00 160.00 140.00 Fn [N] 120.00 100.00 80.00 60.00 40.00 20.00 -1.000 0.00 0.000 1.000 2.000 3.000 4.000 5.000 U [V] measured normal force calculated normal force(Fn = 36.80005 * U + 3.99248) Ill. 36: Characteristic curve high temperature normal force, with room temperature force module Damian Frey Simon Wüest 27 / 69 High-Temperature Tribometer