Download Operating Manual - Molbio Diagnostics
Transcript
Trueprep™ MAG 7.5.6 The message “Step 4: WASH-2” will be displayed on the screen. 7.5.6.1 You will be prompted on the screen with the message “Add 1.0 ml of WASH BUFFER B and Press NEXT”. 7.5.6.2 Add 1.0 ml of Wash Buffer B to the EXT with a 1 ml pipette. 7.5.6.3 Press the Next Button to begin this step, which takes 90 seconds. 7.5.6.4 At the end of this step you will be prompted with the m e s s a g e “ A S P I R AT E LIQUID and Press NEXT”. 7.5.6.5 Aspirate the liquid in the EXT using a 1 ml pipette. Aspirate twice to remove all liquid. 7.5.6.6 Press the Next Button to proceed to the next step. Step 4 : WASH-2 Add 1.0 ml WASH BUFFER B and Press NEXT Step 4 : WASH-2 ---PROCESSING--90 Seconds Remaining Step 4 : WASH-2 ASPIRATE LIQUID and Press NEXT 7.5.7 The message “Step 5: WASH-3” will be displayed on the Screen. 7.5.7.1 You will be prompted on the screen with the message “Add 1.0 ml WASH BUFFER B and Press NEXT”. 7.5.7.2 Add 1.0 ml of Wash Buffer B to the EXT with a 1 ml pipette. 7.5.7.3 Press the Next Button to begin this step, which takes 90 seconds. Step 5 : WASH-3 ADD 1.0 ml WASH BUFFER B and Press NEXT Step 5 : WASH-3 ---PROCESSING--90 Seconds Remaining 9 DIAGNOSTICS