Figure 12. String pot Side impact bracket
Note: A spring is used to keep tension on the string potentiometer cable. When dismounting the string
potentiometer from the rib, gently let the cable be reeled in by the spring. Releasing the cable while it is
pulled out will cause serious damage the string potentiometer.
Side Impacts
1. To use the string potentiometer for side impact, remove the string potentiometer caddy from the
cavity in the thoracic spine. Remove the string potentiometer from its caddy. The string
potentiometer can be used for left and right sided impacts by mounting the potentiometer on the
right of left side.
2. Fasten the string potentiometer on the side impact caddy (048-3103) using the two imperial 2-56
screws. Note that the 2-56 x 3/8 screws are used at the top, while a shorter 2-56 x 5/16 screw is
used at the bottom hole.
3. Attach the caddy to the thoracic spine, using the two M3 x 20 SHCS screws.
4. Pull out the string towards the rib, and put the ball in the socket of the string potentiometer
attachment.
5. Insert the screw though the rib and fasten the string potentiometer attachment to the rib cage
assembly.
6. To remove the string potentiometer, perform steps 2-4 in reverse order and fasten the string
potentiometer attachment to the rib.
4.5
Lumbar Spine
Construction
The lumbar spine consists of
 lumbar spine central molding
 cable assembly
 load cell structural replacement
 screws (M5x12)