Fragment DNA
3
Fragment the DNA using the following settings:
Table 5 Covaris S220 Settings
Setting
350 bp Insert
550 bp Insert
Duty factor
5%
Peak Incident Power
175 W
Cycles per burst
Duration
200
50 seconds
Mode
25 seconds
Frequency sweeping
Temperature
5.5° to 6°C
NOTE
The Covaris M220 settings are optimized for use with the Covaris microTUBE AFA
Fiber Pre-Slit Snap-Cap 6x16mm.
Table 6 Covaris M220 Settings
Setting
350 bp Insert
550 bp Insert
Duty factor
20%
Peak Incident Power
50 W
Cycles per burst
Duration
200
65 seconds
45 seconds
Temperature
20°C
Table 7 Covaris S2 and E210 Settings
Setting
350 bp Insert
Duty cycle
Intensity
10%
5.0
2.0
Cycles per burst
200
Duration
Mode
Displayed Power
Temperature
550 bp Insert
45 seconds
Frequency sweeping
S2—23 W
S2—9 W
E210—14 W
E210—7 W
5.5° to 6°C
4
Centrifuge the plate at 280 × g for 5 seconds.
5
Transfer 50 µl of fragmented DNA from each Covaris tube in the CFP plate to the
corresponding well of the new 0.3 ml PCR plate labeled with the CSP barcode, using a
single channel pipette.
TruSeq DNA PCR-Free Library Prep Guide
13