Anesthetic Gas Measurement
19.7 Anesthetic Gas Accessories
„
Sampling elbow
PRYON P/N: 000.91400
„
Airway adapter for Adult
PRYON P/N: 000.91060
„
DRYLINE Water Trap Adult
ARTEMA P/N:60-13100-00
„
DRYLINE Water Trap Neonate
ARTEMA P/N:60-13200-00
„
DRYLINE Sample Line Adult 1.5m
ARTEMA P/N:60-15100-00
„
DRYLINE Sample Line Adult 2.5m
ARTEMA P/N:60-15200-00
„
DRYLINE Sample Line Neonate 2.5m
ARTEMA P/N:60-15300-00
19.8 Maintenance and cleaning
„
AG module
For detailed cleaning information about “AG Module”, refer to the chapter of
“Maintenance and Cleaning” in this operation manual.
„
Bacteria filter
The bacteria filter is one-off type, i.e., one bacteria filter can only be used by one patient.
„
Sample line
The sample line is one-off type.
„
Gas exhaust outlet
The gas exhaust outlet is reusable. You need to replace it only when it is damaged or
becomes loosely connected. This tube can be cleaned and disinfected.
Cleaning: use cloth moistened with warm soap water to clean the tube. Do not immerse
the tube into the liquid.
Disinfection: use cloth moistened with cool chemical disinfector (ramification mainly
containing aldehyde, ethanol or ramification mainly containing ethanol) to clean the tube.
Do not immerse the tube into the liquid. After cleaning, use wet cloth to wipe off the
disinfector and then use dry cloth to wipe the tube.
„
Watertrap
Replace the watertrap periodically every month or when prompt message appears
requiring the maintenance.
„
Occlusion handling
If the AG module passage is occluded, the screen will display the message “AG
OCCLUSION”. Following are a few examples of occlusion, which you may remove one
by one until this message disappears.