Transcript
Anesthetic Gas Measurement 19.7 Anesthetic Gas Accessories Sampling elbow PRYON P/N: 000.91400 Airway adapter for Adult PRYON P/N: 000.91060 DRYLINE Water Trap Adult ARTEMA P/N:60-13100-00 DRYLINE Water Trap Neonate ARTEMA P/N:60-13200-00 DRYLINE Sample Line Adult 1.5m ARTEMA P/N:60-15100-00 DRYLINE Sample Line Adult 2.5m ARTEMA P/N:60-15200-00 DRYLINE Sample Line Neonate 2.5m ARTEMA P/N:60-15300-00 19.8 Maintenance and cleaning AG module For detailed cleaning information about “AG Module”, refer to the chapter of “Maintenance and Cleaning” in this operation manual. Bacteria filter The bacteria filter is one-off type, i.e., one bacteria filter can only be used by one patient. Sample line The sample line is one-off type. Gas exhaust outlet The gas exhaust outlet is reusable. You need to replace it only when it is damaged or becomes loosely connected. This tube can be cleaned and disinfected. Cleaning: use cloth moistened with warm soap water to clean the tube. Do not immerse the tube into the liquid. Disinfection: use cloth moistened with cool chemical disinfector (ramification mainly containing aldehyde, ethanol or ramification mainly containing ethanol) to clean the tube. Do not immerse the tube into the liquid. After cleaning, use wet cloth to wipe off the disinfector and then use dry cloth to wipe the tube. Watertrap Replace the watertrap periodically every month or when prompt message appears requiring the maintenance. Occlusion handling If the AG module passage is occluded, the screen will display the message “AG OCCLUSION”. Following are a few examples of occlusion, which you may remove one by one until this message disappears. Patient Monitor user’s manual (V 2.2) 19-17