Download Mindray Wato EX-65 Anaesthesia Machine
Transcript
6.9.2 Test the O2 Concentration Monitoring and Alarms NOTE z This test is not required if no O2 sensor is configured. 1. Set the bag/mechanical ventilation switch to the bag 2. Remove the O2 sensor. After two to three minutes, make sure that the sensor measures approximately 21% O2 in room air. 3. Select the [Alarm Setup] shortcut key and then [Ventilator >>]. Set the FiO2 low alarm limit to 50%. 4. Make sure that a low FiO2 alarm occurs. 5. Set the FiO2 low alarm limit back to a value less than the measured FiO2 value and make sure that the alarm cancels. 6. Put the O2 sensor back in the breathing system. 7. Select the [Alarm Setup] shortcut key and then [Ventilator >>]. Set the FiO2 high alarm limit to 50%. 8. Connect the manual bag to the manual bag port. Push the O2 flush button to fill the manual bag. Make sure that the sensor measures approximately 100% O2. 9. Make sure that a high FiO2 alarm occurs. position. 10. Set the FiO2 high alarm limit to 100% and make sure that the alarm cancels. 6.9.3 Test the Low Minute Volume Alarm 1. Make sure that MV alarm is switched on. 2. Select the [Alarm Setup] shortcut key and then [Ventilator >>]. Set the MV low alarm limit to 8.0 L/min. 3. Make sure that a low MV alarm occurs. 4. Select the [Alarm Setup] shortcut key and then [Ventilator >>]. Set the MV low alarm limit to 2.0 L/min. 6-13