Specifications
Wavelength Range
Spectral Bandpass
Wavelength Accuracy
Wavelength Repeatability
Stray Radiant Energy
Photometric Range
Photometric Accuracy
Power Requirements
Dimensions
Weight
Model 1200
325-1000nm
5nm
Better than 2nm
Better than 1nm
<0.5 @ 340 and 400nm
0 to 125%T
0 to 2.0 Abs
0 to 1999C
± 1.0%T
115/230V ±10%, 60/50 Hz
Switchable
408W X 308D X 185H (mm)
15 lbs(6.8kg)
Model 1201
325-1000nm
5nm
Better than 2nm
Better than 1nm
<0.5 @ 340 and 400nm
0 to 125%T
0 to 2.0 Abs
0 to 1999C
± 1.0%T
115/230V ±10%, 60/50 Hz
Switchable
408W X 308D X 180H (mm)
15 lbs(6.8kg)
Installation:
1. After carefully unpacking the contents, check the materials with the packing list
(page 5) to ensure that you have received everything in good condition.
2. Place the instrument in a suitable location away from direct sunlight. In order to
have the best performance from your instrument, keep it as far as possible from any
strong magnetic or electrical fields or any electrical device that may generate highfrequency fields. Set the unit up in an area that is free of dust, corrosive gases and
strong vibrations.
3. Remove any obstructions or materials that could hinder the flow of air under and
around the instrument.
4. Select either 230V or 115V on the voltage selector switch illustrated below, to match
your local mains voltage supply.
5. Turn on your Cole Parmer 1200 and allow it to warm up for 15 minutes before taking
any readings. (Note: picture shows Cole Parmer 1100 series which is similar to 1200
series on the back view. 1200 series has RS-232C Port)
6