Download Otoport Lite Manual Issue 3
Transcript
Otoport Lite 105 User Manual Cavity test 11.2.a Due to mechanical or electrical failure or a faulty probe, the system may itself produce signals during a test that have similar characteristics to an OAE response. The test cavity will not produce any OAE signal so any signal produced in the cavity is the result of a fault. 11.2.b By performing an OAE test in a cavity rather than an ear it is possible to determine whether signal responses recorded in the ear are being generated by a fault. 11.2.c 11.2.d 11.2.e 11.2.f 11.2.g 11.2.h 11.2.i 11.2.j Before starting a cavity test: for the TE model, change the Test Settings Display to Circles; for the DP model, change the stimulus level to 65/55 (see chapter 8 Configuration). Follow the instructions on the previous page for inserting the probe into the test cavity. Start a test on the Otoport as you would for an ordinary ear (see chapter 5 Test ). Follow the Checkfit and Test screen sequences until the test stops. In a cavity the Checkfit screen should indicate a large cavity volume. If the result No Valid OAE is displayed and the Circles are no more than 25 percent full when the test stops, the cavity test has passed. If the result Stopped Too Soon, Too Noisy or Poor Probe Fit is achieved, retest checking that the probe ear piece is firmly inserted into the test cavity and that the noise conditions within the room are acceptable for a test to be conducted. Continue to retest until either No Valid OAE or Pass result is given. If Pass is shown at the end of the test or if the circle for any band is more than ¼ filled, save and retest making sure the ear piece has been firmly pressed into the test cavity. Check the top of the test cavity and ensure it is securely attached to the clear plastic part of the test cavity. If the resources are available, repeat the test with a different test cavity and then with a different probe. This will identify which component is responsible for the problem. Contact your dealer or Otodynamics for further advice. Probe checks 11.2