Download Helena - Handicare
Transcript
Mount secondary rail on traverse rail carrier assembly - Insert the rail over the square nuts on the traverse rail carrier. Tighten with a 17mm open-end wrench (image 7). 7 M10x25 M10x16 M10x25 M10x16 Max 50mm! 7b - 7c NOTE! Ensure that secondary rail overhang does not exceed 50mm! Risk for instability! (Image 7b) - NOTE! Make sure the long bolt (10) is inserted into the secondary rail’s safety hole (image 7c). Finally, mount the following on the secondary rail in the following order: - End stop - Lift unit with carriage - End stop - Screw covers on all visible bolt heads Adjust Helena to the desired height by first loosening the locking knobs on the support leg pairs (one pair at a time) and then carefully rocking the lift stand. Mount the locking knobs in appropriate holes. - NOTE! Maximum height must not be exceeded! The red warning label must not be visible when the desired height has been set. Risk for instability! (Image 8) 8 6 M A N U A L SystemRoMedic TM