Operational Considerations
3
Operational Considerations
3.1
Measuring Principle
The measuring principle of a capacitive dimensional measurement system is
based on the function of an ideal parallel-plate capacitor. The sensor probe
surface and the conductive target surface form the two plate electrodes.
The measurement itself is a measurement of the capacitance between sensor
probe and target surface, which is directly proportional to the change in the
gap.
Fig. 9: Circuitry principle for single-electrode capacitive measurement. For the
nominal range, Cref is 5 pF
The PISeca sensor probes feature a special guard electrode that guarantees
the homogeneity of the electric field by protecting it from outside influences.
Fig. 10: Capacitive sensor working principle. The capacitance C is proportional
to the active area A, Ɛ0 is constant, Ɛr is the dielectric constant of the
material between the plates, generally air