Transcript
Automated Characterization Suite (ACS) Release Notes Changed the note in Diagnostics module. Repaired the spelling error of "SLot." Enabled the prober returning to the home position in single wafer mode. Repaired the mismatching in cassette info. Repaired the CVU locked when several times run system mode test without reset and reconnect KXCI problem. Repaired the Program not responsive or “hang" problem. Enabled the NBTI supporting TAR_IATV(). Enabled the NBTI supporting VTCI_LOG10(). Disabled the General Purpose Instrument to follow matrix Common LO/Independent LO scheme. Added a Qbd log stress value as attribute. Repaired the ACS data disappearing in trial operation problem. Repaired the ACS wafer level plot problem. Added a function to find out Qbd breakdown in ACS and KDAT. Added a breakdown judgment to limit lowest voltage in JRamp. Enabled the Chuck_Force items in 3730 column. Repaired the 237 is not recognize after configure Hardware problem. Repaired the axis title and axis values over lap if more than 1 plot problem. Added an error checking for Groups and associated SMU's. Added a bsweep Python lpt for 4200. Enabled the 2410 over-temperature error displayed in ACS. Repaired the prober button is masked by abort operation problem. Removed the cassette map information in the project. Enabled the writing data to kdf after each device. Known issues The KI2651A instrument model is not supported within an ITM and 2600-based-RTMs. In Windows 7 operating system, you will need to keep the network cable plugged-in if you are using a license file. If the network cable is removed, ACS consider your license file invalid. 16 PA-1008 Rev. H / January 2013