PA-96W User Manual: Rev.1-11/04
4.4.8
Luciferase assay at 24-48 hours
• Add 100 µl lysis buffer with DTT to each well
• Mix each well by pipetting up and down 12 times
• Add 75 µl lysed cell suspension to wells of a white assay plate for each
sample
• Standard curve (if used)
ƒ Remove luciferase standard from –70o C freezer (1 mg/ml)
ƒ Add 400 lysing solution to each of 9 tubes
ƒ Add 1 ml lysing solution to 1 tube (Tube A)
ƒ Add 10 µl luciferase to tube A, vortex (1:100 dilution)
ƒ Transfer 10 µl from tube A to the first of the 9 tubes (1:41 dilution),
vortex
ƒ Transfer 200 µl to the second of 9 tubes (1:3 dilution)
ƒ Serially transfer 200 µl to the remaining tubes
ƒ (use last 8 tubes, discard first two)
ƒ Add 75 µl standard to duplicate wells of the white plate
• Place 50 µl reagent A in each well
• Place reagent B in Luminometer injector and prime
• Inject 50 µl
• Read for 0.5 seconds per well
Demonstration of K562 Cell Transfection Using
96 W System
2000 V/cm
120000
Luciferase RLU
100000
100 us
200 us
400 us
800 us
80000
60000
40000
20000
0
1 Pulse
3 Pulses
6 Pulses
Number of Pulses versus Pulse Width
Figure 4-1 Example K562 Protocol results