when the clock of each instrument reaches the time
corresponding to the beginning of the next segment, the
measurement in progress will be stopped and stored, and a
new measurement will be started.
Other Measurement Times
When values of measurement other than those listed above
are used, the clock time at which the first measurement is cut
short is based solely on the units of the measurement time;
seconds, minutes or hours. For example, suppose the
measurement time is set to three minutes. Based on the unit
of minutes, the time intervals per hour are xx:00:00,
xx:01:00, xx:02:00, etc. If the measurement were begun at
08:14:23, the first measurement would be cut short at the
next measurement interval, so subsequent measurements
would be started at 08:15:00, 08:18:00, 08:21:00, etc.
Timed Stop Mode
The Time Stop Mode with Measurement History has a
feature not included for the other run modes; the ability to
automatically measure and store a user-defined number of
records, then stop. Subsequent runs, each manually initiated,
will produce the same number of stored measurements.
Measurement Counter
When the Measurement History has been enabled for the
Time Stop Mode, the Measurement Counter field will
appear as shown in Figure 10-3.
FIGURE 10-3 Measurement Counter
Measurement Counter is used to set the number of
measurements to be made during each manually initiated
measurement sequence.
Note that the measurement counter
will be set to one and the selection
field disabled when the measurement
time has been set to less than one
minute.
LxT Technical Reference Manual
To enter the a value into the Measurement Counter field,
time, highlight the Measurement Counter data field and
press the
key. Enter the desired value and press the
key to conclude the process.
5
5
Timed Stop Mode
10-3