Download 1083-150 C2 PPMS Cry..
Transcript
Section 4.2 Troubleshooting 4.2.2 Chapter 4 Troubleshooting and Maintenance Exposure to Atmosphere If the sorption pump has been accidentally exposed to the atmosphere, it is probably saturated with frozen air and might require an out-of-dewar regeneration (Section 4.3) before it operates properly. 4.2.3 Flapper Valve Fails to Open The color-coded motor-housing indicator indicates the position of the flapper valve (see Figure 3-2). When the flapper valve opens, the indicator should show red. If it continues to show black, or both red and black, when attempting high vacuum, investigate whether one or more of the following situations has occurred: ○ The spring has failed. Remove the cover panel and inspect the spring. ○ The cam is slipping. ○ The flapper valve has frozen shut. If the flapper valve fails to open, an error message will be generated in PPMS MultiVu. If this error message appears but the flapper valve indicator displays the red stripe (meaning that the flapper valve is actually open), there is a problem with either the wiring or the limit switch. ○ 4.2.4 The motor has failed. Flapper Valve Fails to Close The color-coded motor-housing indicator indicates the position of the flapper valve (see Figure 3-2). When the flapper valve closes, the indicator should show black. If it continues to show red, or both red and black, when terminating high vacuum, investigate whether one or more of the following situations has occurred: ○ The spring has failed. Remove the cover panel and inspect the spring. ○ The cam is slipping. ○ The flapper valve has jammed. If the flapper valve fails to close, an error message will be generated in PPMS MultiVu. If this error message appears but the flapper valve indicator displays the black stripe (meaning that the flapper valve is actually closed), there is a problem with either the wiring or the limit switch. 4-2 ○ The micro switch has failed. ○ The motor has failed. PPMS Cryopump User’s Manual, Rev. C-2 June 2005 Quantum Design