Download For Reference Purpose Only!
Transcript
CircuLex Human ISG15 ELISA Kit User’s Manual For Research Use Only, Not for use in diagnostic procedures On ly! TM Detailed Protocol The CycLex Research Product CircuLex Human ISG15 ELISA Kit is provided with removable strips of wells so the assay can be carried out on separate occasions using only the number of strips required for the particular determination. Since experimental conditions may vary, an aliquot of the human ISG15 Standard within the kit, should be included in each assay as a calibrator. Disposable pipette tips and reagent troughs should be used for all liquid transfers to avoid cross-contamination of reagents or samples. Preparation of Working Solutions rp os e All reagents need to be brought to room temperature prior to the assay. Assay reagents are supplied ready-to-use, with the exception of 10X Wash Buffer and Human ISG15 Standard. 1. Prepare a working solution of Wash Buffer by adding 100 mL of the 10X Wash Buffer to 900 mL of deionized (distilled) water (ddH2O). Mix well. Store at 4°C for two weeks or -20°C for long-term storage. 2. Reconstitute Human ISG15 Standard with 1.0 mL of ddH2O. The concentration of the human ISG15 in vial should be 40 ng/mL, which is referred as a Master Standard of human ISG15. Pu Prepare Standard Solutions as follows: Use the Master Standard to produce a dilution series (below). Mix each tube thoroughly before the next transfer. The 12 ng/mL standard (Std.1) serves as the highest standard. The Dilution Buffer serves as the zero standard (Blank). Volume of Standard 180 µL of Master Standard 300 µL of Std. 1 (12.0 ng/mL) 300 µL of Std. 2 (6.0 ng/mL) 300 µL of Std. 3 (3.0 ng/mL) 300 µL of Std. 4 (1.5 ng/mL) 300 µL of Std. 5 (0.75 ng/mL) 300 µL of Std. 6 (0.375 ng/mL) en ce Std.1 Std.2 Std.3 Std.4 Std.5 Std.6 Std.7 Blank - Dilution Buffer 420 µL 300 µL 300 µL 300 µL 300 µL 300 µL 300 µL 300 µL Concentration 12.0 ng/mL 6.0 ng/mL 3.0ng/mL 1.5 ng/mL 0.75 ng/mL 0.375 ng/mL 0.188 ng/mL 0 ng/mL er Note: Do not use a Repeating pipette. Change tips for every dilution. Wet tip with Dilution Buffer before dispensing. Unused portions of Master Standard should be aliquoted and stored at below -70°C immediately. Avoid multiple freeze and thaw cycles. ef Sample Preparation • Optimal dilution of cell conditioned medium for measurement of ISG15 depends on cell number and IFN-does; require neat to appropriate dilution. rR • Serum and plasma samples require a 10-40 fold dilution depending on ISG15 concentration in sample. Fo Cat#: CY-8085 7 Version#: 120710