PRELIMINARY P2-80-12K40 User’s Manual 48 To perform PRNU calibration to a camera calculated peak value, use the command: Purpose: Performs PRNU calibration and eliminates the difference in responsivity between the most and least sensitive pixel creating a uniform response to light. Using this command, you do not have to provide a calibration target. Syntax: ccp [i] Syntax Elements: [i] When the optional parameter is provided, each tap’s analog gain so that 8-13% of pixels within a tap are above the value specified in the target value. This command sets the analog gain first to i before calculating the PRNU coefficients range from 256 to 1004DN. If value is left blank, current gain values are used. • Notes: • • Example: This command performs the same function as the cpa 1 i command but does not force you to enter a target value. Perform all analog and digital adjustments before calibrating PRNU. Calibrate FPN before calibrating PRNU. ccp Figure 15: PRNU Calibration (ROI=1-12288) (ccp 800 or cpa 1 800) Video Gain clipped to 3.99 1023DN 800DN After PRNU correction 511DN Before PRNU correction 400DN 200DN 100DN 0 12288 Tap 1 odd Tap 2 even Tap 5 odd Tap 6 even Tap 3 odd Tap 4 even Tap 7 odd Tap 8 even Pixel Number DALSA 03-32-10128-03