7 Performance Tests and Calibration Output calibration 1 Set the slide switch to the M/S position. 2 Allow the instrument to warm up for 10 minutes before performing calibration. 3 To enter the calibration mode, press and for more than one second. The primary display will indicate “CHEEP”. 4 Press to enter the input calibration mode. 5 Turn the rotary switch to any one of the “Current Input/Voltage Output” positions, and press SHIFT for more than one second to enter the output calibration mode. CAL-0 & CAL-1 In the output calibration mode, the primary and secondary display will show “CAL- 0” and “- rdy- ” respectively. Connect the output terminals to a multimeter (refer to Table 7- 1 on page 131 for the recommended test equipment). • CAL-0: 1 Press OUTPUT. The primary and secondary displays show “CAL- 0” and “00000” respectively. 2 Wait until the instrument reading becomes stable, then record the value. • CAL-1: 1 Press MODE. The primary and secondary displays show “CAL- 1” and “- rdy” respectively. 2 Press OUTPUT. The primary and secondary displays show “CAL- 1” and “00000” respectively. 3 Press or to adjust the output voltage until the reading on the meter is the same as the CAL-0 value recorded above. 4 Press MODE to finish the CAL-0 and CAL-1 calibration. 144 U1401B User’s and Service Guide