Download CardEX - Midmark

Transcript
Charge voltage (standard) = (16.8-0.1V)
Cycle life ≥ 300 times
Recording
HR
Recognition
ECG Unit
Power Consumption:
35VA (max)
Fuse:
T400mA250V Ø5×20 / T200mA 250V Ø5×20
Recorder:
Thermal dot-matrix printer
Recorder Paper:
Folded thermal paper, 80mm×70mm×200pages
Rolled thermal paper, 80mm×20m
Effective Width:
72mm
Paper Speed:
5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s,
50mm/s ±3%
Accuracy of data:
±5% (x-axis), ±5%(y-axis)
Technique:
Peak-peak detection
HR Range:
30 BPM ~300 BPM
Accuracy:
±1BPM
Leads:
7 standard leads
Acquisition Mode:
simultaneous 7 leads
A/D Resolution:
12bits
Time Constant:
≥3.2s
Frequency Response:
0.05Hz ~ 150Hz (-3dB)
Sensitivity:
2.5, 5, 10, 20 (mm/mV)
Input Impedance:
≥50MΩ(10Hz)
Input Circuit Current:
≤0.05μA
Input Voltage Range:
<±5 mVpp
Calibration Voltage:
1mV±3%
DC Offset Voltage:
±500mV
Noise:
≤12.5μVp-p
Multi-channel Crosstalk
≤0.5mm
Filter:
AC Filter: On/Off
( )
- 47 -