Download Medical electrical equipment
Transcript
VICTRONIC TECHNOLOGY CORPORATION IEC 60601-1 Clause 7.9.3.2 Requirement + Test Result - Remark Verdict MANUFACTURER’S optional requirements for minimum qualifications of SERVICE PERSONNEL documented in technical description N/A The technical description contains the following required information N/A –type and full rating of fuses used in SUPPLY MAINS external to PERMANENTLY INSTALLED ME EQUIPMENT, when type and rating of fuses are not apparent from information on RATED current and mode of operation of ME EQUIPMENT .......... : Not permanently installed ME equipment N/A – a statement for ME EQUIPMENT with a nonDETACHABLE POWER SUPPLY CORD if POWER SUPPLY CORD is replaceable by SERVICE PERSONNEL, and Detachable power supply cord used. N/A No such instruction is required N/A if so, instructions for correct connection and anchoring to ensure compliance with 8.11.3 – instructions for correct replacement of interchangeable or detachable parts specified by MANUFACTURER as replaceable by SERVICE PERSONNEL, and – warnings identifying nature of HAZARD when replacement of a component could result in an unacceptable RISK, and when replaceable by SERVICE PERSONNEL all information necessary to safely replace the component N/A 7.9.3.3 Technical description indicates, MANUFACTURER See User Manual: “Safety & will provide circuit diagrams, component part Warranty” lists, descriptions, calibration instructions to assist to SERVICE PERSONNEL in parts repair Pass 7.9.3.4 Means used to comply with requirements of Isolation power supply 8.11.1 clearly identified in technical description provided as source of electricity Pass 8 PROTECTION AGAINST ELECTRICAL HAZARDS FROM ME EQUIPMENT Pass 8.1 Limits specified in Clause 8.4 not exceeded for See appended RM Results ACCESSIBLE PARTS and APPLIED PARTS in NORMAL Table 8.1 b(1), 8.1 b(2) and or SINGLE FAULT CONDITIONS 8.1 b(3) Pass NORMAL CONDITION considered as simultaneous occurrence of situations identified in 8.1a) Pass SINGLE FAULT CONDITION considered to include See appended RM Results the occurrences as specified in Clause 8.1b) ... : Table 8.1 b(1), 8.1 b(2) and 8.1 b(3) Pass ACCESSIBLE PARTS determined according to 5.9 Pass LEAKAGE CURRENTS measured according to 8.7 Pass Requirements related to power sources Pass 8.2 Test Report No.: 1403042002 60601-1TRF_A Page 30 of 197 Issued: 25 July, 2014