Download Medical electrical equipment
Transcript
VICTRONIC TECHNOLOGY CORPORATION IEC 60601-1 Clause Requirement + Test Result - Remark RATED input for one or more RATED voltage ranges provided for upper and lower limits of the range or ranges when the range(s) is/are greater than ± 10 % of the mean value of specified range (A, VA,W) .....................................: For Adapter: AC 1.2-0.63A For Monitor: DC 4.2 A Input at mean value of range marked when range limits do not differ by more than 10 % from mean value (A, VA, W) .................................: No such marking N/A Marking includes long-time and most relevant No such marking momentary volt-ampere ratings when provided, each plainly identified and indicated in ACCOMPANYING DOCUMENTS (VA) .............................: N/A Marked input of ME EQUIPMENT provided with means for connection of supply conductors of other electrical equipment includes RATED and marked output of such means (A, VA, W) ..........: Verdict Monitor: 12 V dc, 4.2 A Power Adapter: 100-240 V ac, 60/50 Hz, 1.2-0.63 A 7.2.8 Output connectors 7.2.8.1 See 16.9.2.1 b) for MULTIPLE SOCKET-OUTLETS integral with ME EQUIPMENT 7.2.8.2 Output connectors are marked, except for Single output for external MULTIPLE SOCKET-OUTLETS or connectors intended adapter and the rating for specified ACCESSORIES or equipment marked on the label. Pass Pass Pass No MULTIPLE SOCKET-OUTLETS N/A Pass Rated Voltage (V), Rated Current (A) ................ : — Rated Power (W), Output Frequency (Hz) ........ : — 7.2.9 ME EQUIPMENT or its parts marked with the IP The equipment classified environmental Code per IEC 60529 according to IPX0 need not be marked. classification in 6.3 (Table D.3, Code 2)..............: N/A 7.2.10 Degrees of protection against electric shock as classified in 6.2 for all APPLIED PARTS marked with relevant symbols as follows (not applied to parts identified according to 4.6): N/A TYPE B APPLIED PARTS with symbol 19 of Table D.1 (IEC 60417-5840, DB: 2002-10), not applied in such a way as to give the impression of being inscribed within a square in order to distinguish it from symbol IEC 60417-5333 ............................: Pass TYPE BF APPLIED PARTS with symbol 20 of Table D.1 (IEC 60417-5333, DB: 2002-10) ......................: N/A TYPE CF APPLIED PARTS with symbol 21 of Table D.1 (IEC 60417-5335, DB: 2002-10) ......................: N/A DEFIBRILLATION-PROOF APPLIED PARTS marked with symbols 25-27 of Table D.1 (IEC 60417-5841, IEC 60417-5334, or IEC 60417-5336, all DB: 2002-10): N/A Test Report No.: 1403042002 60601-1TRF_A Page 19 of 197 Issued: 25 July, 2014