Download Medical electrical equipment

Transcript
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
RATED input for one or more RATED voltage
ranges provided for upper and lower limits of
the range or ranges when the range(s) is/are
greater than ± 10 % of the mean value of
specified range (A, VA,W) .....................................:
For Adapter: AC 1.2-0.63A
For Monitor: DC 4.2 A
Input at mean value of range marked when
range limits do not differ by more than 10 %
from mean value (A, VA, W) .................................:
No such marking
N/A
Marking includes long-time and most relevant
No such marking
momentary volt-ampere ratings when provided,
each plainly identified and indicated in
ACCOMPANYING DOCUMENTS (VA) .............................:
N/A
Marked input of ME EQUIPMENT provided with
means for connection of supply conductors of
other electrical equipment includes RATED and
marked output of such means (A, VA, W) ..........:
Verdict
Monitor: 12 V dc, 4.2 A
Power Adapter: 100-240 V
ac, 60/50 Hz, 1.2-0.63 A
7.2.8
Output connectors
7.2.8.1
See 16.9.2.1 b) for MULTIPLE SOCKET-OUTLETS
integral with ME EQUIPMENT
7.2.8.2
Output connectors are marked, except for
Single output for external
MULTIPLE SOCKET-OUTLETS or connectors intended adapter and the rating
for specified ACCESSORIES or equipment
marked on the label.
Pass
Pass
Pass
No MULTIPLE SOCKET-OUTLETS
N/A
Pass
Rated Voltage (V), Rated Current (A) ................ :
—
Rated Power (W), Output Frequency (Hz) ........ :
—
7.2.9
ME EQUIPMENT or its parts marked with the IP
The equipment classified
environmental Code per IEC 60529 according to IPX0 need not be marked.
classification in 6.3 (Table D.3, Code 2)..............:
N/A
7.2.10
Degrees of protection against electric shock as
classified in 6.2 for all APPLIED PARTS marked
with relevant symbols as follows (not applied to
parts identified according to 4.6):
N/A
TYPE B APPLIED PARTS with symbol 19 of Table D.1
(IEC 60417-5840, DB: 2002-10), not applied in
such a way as to give the impression of being
inscribed within a square in order to distinguish
it from symbol IEC 60417-5333 ............................:
Pass
TYPE BF APPLIED PARTS with symbol 20 of Table
D.1 (IEC 60417-5333, DB: 2002-10) ......................:
N/A
TYPE CF APPLIED PARTS with symbol 21 of Table
D.1 (IEC 60417-5335, DB: 2002-10) ......................:
N/A
DEFIBRILLATION-PROOF APPLIED PARTS marked with
symbols 25-27 of Table D.1 (IEC 60417-5841, IEC
60417-5334, or IEC 60417-5336, all DB: 2002-10):
N/A
Test Report No.: 1403042002
60601-1TRF_A
Page 19 of 197
Issued: 25 July, 2014