Download Lake Shore Model 475 Gaussmeter Manual
Transcript
Lake Shore Model 475 Gaussmeter User’s Manual Some probes also feature temperature compensation. Turning Temperature Compensation OFF causes the Model 475 to ignore this data. To configure the Field Compensation setting, press the Probe key and wait until the Probe Serial number display disappears and the following screen appears as a prompt for field compensation. Select With °® Field Comp Off Use the s or t key to select Off or On. Press Enter to accept the new selection and continue to the next setting screen. Press Escape to cancel the selection and return to the normal display. The next probe setting screen appears as a prompt for changing the temperature compensation setting. Select With °® Temp Comp Off Use the s or t key to select Off or On. Press Enter to accept the new selection and return to the normal display. Press Escape to cancel the selection and return to the normal display. If the probe does not include temperature compensation, this setting is ignored. 5.1.4 Extension Cable The complex nature of Hall-effect measurements makes it necessary to match extension cables to the probe when longer cables are needed. Keeping probes and their extensions from getting mixed up can become a problem when more than one probe is in use. The Model 475 alleviates some of the difficulty by allowing users to match probes to extensions in the field. Stored information can be viewed on the front panel and read over computer interface to ensure proper mating. NOTE: Calibrated probe and extension cable pairs may be purchased from Lake Shore and are recommended for the best accuracy. Extension cables with no calibration may also be used. The Model 475 can then be used to match a probe to the un-calibrated extension cable. To match a probe with an extension cable, verify that the probe is connected to the instrument. To begin the matching process, press and hold the Probe key for approximately 4 seconds. The following screen appears as a prompt for selecting the programming method. Select With °® MPEC Program Use the s or t key to select from MPEC program (extension cables) or MCBL Program (User programmable cables, refer to Paragraph 5.2.1). Press Enter to accept the new selection and continue the matching process. Press Escape to cancel the selection and return to the normal display. The next probe matching screen appears as a prompt to begin the copying process. Attach MPEC cable Enter to Continue 5-2 Advanced Operation