Download MX100 Data Acquisition Unit User`s Manual
Transcript
1.3 Functions of the 4-CH, High-Speed Universal Input Module This module is equipped with four inputs and measures DC voltage, TC, RTD, and digital input (DI) at a minimum measurement interval of 10 ms (the minimum interval is 50 ms if the input type is thermocouple or RTD). Terminal cover Input terminal (clamp terminal) Input Type and Measurement Range The following measurements are possible. Input DC voltage Thermocouple RTD (Measurement current:1 mA) RTD (Measurement current: 2 mA) DI 1-10 Measurement Range Type 20 mV 60 mV 200 mV 2V 6V 20 V 100 V R S B K E J T L U N W KPvsAu7Fe Pt100 JPt100 Pt100 (high resolution) JPt100 (high resolution) Ni100 SAMA Ni100 DIN Ni120 Pt100 JPt100 Pt100 (high resolution) JPt100 (high resolution) Pt50 Cu10 GE Cu10 L&N Cu10 WEED Cu10 BAILEY J263B Level Contact input Rated Measurement Range –20.000 to 20.000mV –60.00 to 60.00mV –200.00 to 200.00mV –2.0000 to 2.0000V –6.000 to 6.000V –20.000 to 20.000V –100.00 to 100.00V 0.0 to 1760.0°C 0.0 to 1820.0°C –200.0 to 1370.0°C –200.0 to 800.0°C –200.0 to 1100.0°C –200.0 to 400.0°C –200.0 to 900.0°C –200.0 to 400.0°C 0.0 to 1300.0°C 0.0 to 2315.0°C 0.0 to 300.0K –200.0 to 600.0°C –200.0 to 550.0°C –140.00 to 150.00°C –140.00 to 150.00°C –200.0 to 250.0°C –60.0 to 180.0°C –70.0 to 200.0°C –200.0 to 250.0°C –200.0 to 250.0°C –140.00 to 150.00°C –140.00 to 150.00°C –200.0 to 550.0°C –200.0 to 300.0°C –200.0 to 300.0°C –200.0 to 300.0°C –200.0 to 300.0°C 0.0 to 300.0K Vth = 2.4V ON: 100 Ω or less, OFF: 10 kΩ or more IM MX100-01E