Technical Information Use 17 The speed value can be changed from 28 to 42 KRPM. Different values are rejected. The speed adj. Sets the LOW SPEED freq. to 2/3. NOTE The speed can be adjusted only when the pump is stopped. Starting the Pump If the forepump and vent device are not operated by the controller, close the vent valve and switch on the forepump. NOTE Before starting the pump the cooling mode must be selected. NOTE With the FRONT panel operation selected, the REMOTE and RS 232 operations are inoperative; conversely, the CYCLE NUMBER and PUMP CURRENT pushbuttons are always active, even when the operating mode selected is REMOTE or RS 232 / 422 / 485. Press the START push-button (or use the remote or RS 232 start signal), and the display shows: S O F T [ - - S T A R T - - - - - - O N - - ] Figure 40 Turbo-V 700HT Controller User Manual / 87-900-877-01 (F) 289/304