Slowly move the flow cell lever to position 0 to disengage the vacuum seal and release
the flow cell.
Figure 21 Flow Cell Lever in Position 0
4
Lift the used flow cell from the flow cell holder.
Clean the Flow Cell Holder
1
Put on a new pair of powder-free latex gloves.
2
Using an alcohol wipe or a lint-free tissue moistened with ethanol or isopropanol,
carefully wipe the surface of the flow cell holder until it is clean.
CAUTION
Do not allow alcohol to drip into the vacuum holes or around the manifolds. Use a low-lint
lab tissue to dry the stage, if necessary.
HiSeq 1000 System User Guide
51
Load a Flow Cell
3