Download HID EVOlution™ - qPCR/STR Setup System Getting Started Guide

Transcript
A
Appendix A Troubleshooting
Table 12 Troubleshooting qPCR/STR PCR results (continued)
Observed problem
Possible reason
Suggested solution
The sample IPC CT is
higher than the IPC CT of
the no template
quantitation control (NTC)
or of the quantitation
standards (for example, if
the sample IPC CT is
approximately two CT
greater than the NTC IPC
CT or the CT of the
standards).
or
Weak amplification (high
CT value and low ΔRn
value) of the human
and/or male-specific
targets and no or weak
amplification of the IPC.
• The DNA concentration is above
25 ng/µL.
• Potential presence of PCR
inhibitors.
• If the DNA concentration is over 25 ng/µL, dilute
the DNA extract, then requantify the sample.
• If the DNA concentrate is below 25 ng/µL, or if
the diluted DNA eluate still produces high IPC CT
compared to the NTC or quantitation standards,
repurify the sample.
CE signal is too high
Sample concentration is greater
than the reported quantification
result.
72
IMPORTANT! Repurification may result in the
loss of additional DNA. Consider proceeding to
amplification with a kit such as the AmpFlSTR®
MiniFiler™ PCR Amplification kit, which is
designed to obtain STR profiles from
compromised samples such as those which may
be inhibited and/or degraded.
• Before the next run:
– To ensure correct pipetting, clean and finger
tighten the DiTi cones, diluter valves, and
syringe fittings. Refer to the Tecan HID
EVOlution™ - qPCR/STR Setup System
Application Manual.
– Confirm that you have sufficient system liquid,
then flush the system and check for air
bubbles and leaks. Refer to the Tecan HID
EVOlution™ - qPCR/STR Setup System
Application Manual.
• Requantify (see IMPORTANT above).
• Before the next run:
– To ensure correct pipetting, clean and finger
tighten the DiTi cones, diluter valves, and
syringe fittings. Refer to the Tecan HID
EVOlution™ - qPCR/STR Setup System
Application Manual.
– Confirm that you have sufficient system liquid,
then flush the system and check for air
bubbles and leaks. Refer to the Tecan HID
EVOlution™ - qPCR/STR Setup System
Application Manual.
HID EVOlution™ - qPCR/STR Setup System Getting Started Guide