Download RF Lock-In Amplifier - Frederick Seitz Materials Research Laboratory
Transcript
Scan - Rel 3-39 Scan and Rel Example In this example we will make transfer function measurements of a device–under–test, or DUT, using the SR844 REF OUT signal in Internal Reference mode. For this example, we assume that the DUT attenuates the REF OUT signal. We will make measurements at several frequencies using the Scan Mode. We will use Rel Mode to store the measured values when the DUT is by-passed (using STORE R[dBm]θ). First, we set the frequency scan parameters to select the scan frequencies. Then we step through the scan with the DUT by-passed, storing Rel Configurations and Values at each frequency (using STORE R[dBm]θ). This measures the DUT input. Next we pass the signal through the DUT and step through the frequency scan again. This measures the transfer response of the DUT directly in dB and degrees. If the DUT amplifies the signal, we would need to make the scan with the DUT in the signal path first, storing Rel Values and Configurations. The second scan by-passes the DUT and measures the transfer response (negate both the dB and phase readings). The order is determined by which configuration (through the DUT or by-pass the DUT) has the larger output signal. This determines the required measurement configuration at each frequency. 1 Connect the REF OUT signal (appropriately attenuated and terminated) to the input of the DUT. The DUT output should be connected to the SIGNAL IN of the SR844. Now by-pass the DUT, leave as much of the apparatus in place as possible. It is important to use the same cabling for both setups (by-pass and through the DUT) in order to preserve phase delays through the system. 2 Set the frequency scan parameters using the Scan Set key and knob as described previously in the section Using Frequency Scans. Press CLEAR ALL [Shift– Rel Mode] to discard any previously stored Rel Values. 3 Press Start/Step once to begin/continue the frequency scan. 4 Adjust the instrument configuration to measure the signal appropriately. Set the displays to show R[dBm] on CH1 and θ on CH2. This measures the DUT input. 5 Press STORE R[dBm]θ. This saves the signal phase and the measured value of R[dBm] as well as the measurement configuration to be used in the next scan. 6 Repeat steps 3–5 until the Scan is done. 7 Now pass the signal through the DUT. 8 Press Start/Step once more to turn Scan Mode off at the Stop frequency. 9 Press Rel Mode to turn on REL MODE and use the stored Rel Values. The next scan will display R[dBm] and phase relative to the original scan taken with the DUT by-passed. This is exactly the transfer function we started out to measure. 10 Press Start/Step once to begin/continue the frequency scan. 11 When the measurement is stable, note the readings. In dB and degrees, these give the value of the transfer function of the DUT at the current frequency. 12 Repeat steps 10–11 until the Scan is done. SR844 RF Lock-In Amplifier