Download TRobot user manual
Transcript
systatus BBBBBBBB HHHHu system status bit 0 - low voltage bit 1 - block error (controller,lid,cooler) bit 2 bit 3 bit 4 - remote control error bit 5 - memory error bit 6 - software error 0 ... FF tempform C HHHHu temperature format of 'A' tempinc ±DD.DD HHHHs * set temp. increment/C 0 .. ±20.00 timeinc DDD HHHHu set time increment /s 0 .. 240 timeoutlen DDD HHHHu length of timeout 0 .. 255 (*64ms) DDD HHHHu waiting time of repeat function 0 .. 254 (*64ms) if crc16-check is selected waittime 6.10.1 Examples Query of block temperature :b 1;l 6 Bytes in mnemonics :BLCK 1;TEMP 12 Bytes 6.11 Error codes Error codes consist of three decimal numbers. The first sign determines the error group.Variable numerical values of several formats in this description are represented by format signs, beginning with ”%”. Examples: %4x ; hexadecimal, 4 digit %2d ; integer, 2 digit %5f1 ; decimal value, signed, 5 digit altogether, consisting of 3 integer digits, decimal ;point and 1 post-decimal digit. All error texts end with “!”. See also chapter 6.2.3.2. Special error code: !? checksum error 6.11.1 Error Group 0xx: System Errors 001 002 003 004 005 006 56 remote control off! Error %4x %4x %4x %4x %4x %4x %4x wait: too cold! cooler too hot! no printer! no heated lid! ; remote control not activated ; software error, system halted ; cooler too cold ; cooler too hot ; no printer available ; heated lid not connected Instruction Manual TRobot 10/2009