Download MultiRAE User`s Manual
Transcript
MAINTENANCE 5. Plug a new sensor into the empty sensor socket. Make sure that black line on the sensor label is lined up with the white marker on the PCB and the sensor pins are aligned with the socket before pushing down the sensor. 6. Replace the gas piping plate and tighten the four screws to hold down the sensors. Re-connect the battery pack. Replace the monitor cover. 7. Turn on the power of the MultiRAE PLUS. The microprocessor will automatically recognize the sensors installed and configure the monitor accordingly. Special Bias Voltage for Toxic Sensors: The NO and NH3 toxic sensors require a special 300 mV bias voltage to operate. Only the Tox1 sensor position provides such a special bias voltage. Therefore, the NO or NH3 toxic sensor must be plugged into the Tox1 sensor socket in MultiRAE PLUS. In addition, there is a dip switch or pin jumper located on the analog PCB next to the PID sensor. When biased toxic sensor is plugged into the Tox1 position, it is necessary to turn the dip switch or move the jumper to the “ON” position to enable the bias voltage. During power on sequence, the microprocessor will check the sensor ID and the bias voltage. If the NO or NH3 sensor is plugged into the wrong toxic sensor socket or the bias voltage is not switched on, an error message will be displayed. 7-6