To edit a particular Size Standard peak: Step 7 (continued) Action If you choose Add or Change, move the cursor to the right to open the Select Size sub-menu. Choose the desired height value from this menu. 8 Check the sizing quality by selecting Check Sizing Quality under the Tools menu. If the peaks are sized correctly, but the quality score is still below passing, click the Override SQ button to set the SQ to 1.0. 9 Click Apply if you are going to edit a second size standard peak, or click OK if you are finished applying your changes to the edited peak. Using the Size Standard Settings 10-17 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com