Download Cycle end - PTB Sales
Transcript
C 110 Cycle end Roughing timer If Cycle end is activated (automatic), user could activate or not the roughing timer. Roughing timer determines maximum authorized roughing time before the tested part is considered as bad. To consider the tested part as bad, it is necessary to activate the Memo function. ➪ see below ➪ see "Measure" paragraph. Activate the roughing timer F1 F3 F3 + F3 F1 F3 F1 F4 F3 Adjust roughing timer. F1 F3 CYCLE END:automatic *****ROUGHING***** OVERLAP TIMER: no TIMER (min:s):00:09 F2 CYCLE END:automatic *****ROUGHING***** OVERLAP TIMER: yes TIMER (min:s):00:09 F2 CYCLE END:automatic *****ROUGHING***** OVERLAP TIMER: yes TIMER (min:s):00:09 F2 CYCLE END:automatic *****ROUGHING***** OVERLAP TIMER: yes TIMER (min:s):00:05 F2 Adjustment of a value F4 F1 F3 F4 F4 F4 F4 C 20 CYCLE END:automatic *****ROUGHING***** OVERLAP TIMER: yes TIMER (min:s):00:05 F2 F4 Adjust measure timer: see “Measure” paragraph. Please follow the procedure with “Memo function” paragraph. Measure If Cycle end is activated (automatic), user should adjust the measure timer i.e. the cycle end. The measure timer corresponds to the time allowed to the leak detector to remain in test. When it is reached, the measure flashes. 2/3 Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual GB 00267 - Edition 03 - September 01 Activate the roughing timer ■ Not activate the roughing timer ■