Download 1260 Infinity High Performance Micro Autosampler
Transcript
11 Hardware Information Setting the 8-bit Configuration Switch (without On-board) LAN Setting the 8-bit Configuration Switch (without On-board) LAN The 8- bit configuration switch is located at the rear of the module. This module does not have its own on- board LAN interface. It can be controlled through the LAN interface of another module, and a CAN connection to that module. Figure 31 Configuration switch (settings depend on configured mode) All modules without on- board LAN: • default should be ALL DIPS DOWN (= best settings) • Bootp mode for LAN and • 19200 baud, 8 data bit / 1 stop bit with no parity for RS- 232 • DIP 1 DOWN and DIP 2 UP allows special RS- 232 settings • for boot/test modes DIPS 1+2 must be UP plus required mode NOTE For normal operation use the default (best) settings. Switch settings provide configuration parameters for serial communication protocol and instrument specific initialization procedures. NOTE 190 With the introduction of the Agilent 1260 Infinity, all GPIB interfaces have been removed. The preferred communication is LAN. 1260 Infinity Autosampler User Manual