Transcript
110 tasks using the software to control the CT scanner will be described here: capturing an x-ray image, running a CT scan, and reconstructing a CT image. Capturing an X-Ray Image Capturing an x-ray image is similar to an x-ray technician at a hospital performing an x-ray on a patient’s broken bone. The resulting image is a view of an object’s internal features, but seen as a two-dimensional image. Internal features may be identified, but their exact location is unknown. The steps to capture an x-ray image using the MSU CT scanner are as follows: 1. Turn on the x-ray console and warm it up, and turn on the power to the CT cabinet. 2. Start the “9621” software from the Windows desktop to control the CT equipment. 3. Click the “Home” button for the table’s vertical position to re-zero the table’s location. 4. Move the table to an appropriate height so that the feature you wish to view is level with the x-ray converting screen. You can view the table’s movement by opening the cabinet and pulling the small switch near the bottom of the center post. Be sure to close the door before turning on the x-ray. 5. Remove the lead collimators from the x-ray and the converting screen to capture a full-screen x-ray image. If the collimators are left on, the image will contain only a thin slice of the object.