Download CE EMC Test Report
Transcript
10 Voltage Fluctuations and Flicker Measurement 10.1 Limits Test item Pst Plt Limit 1.0 0.65 Tmax (ms) 500 dmax (%) 4 dc (%) 3.3 Note Pst: short-term flicker severity. Plt: long-term flicker severity. Tmax: maximum time duration during the observation period that the voltage deviation d(t) exceeds the limit for dc . dmax: maximum absolute voltage change during an observation period. dc: maximum steady state voltage change during an observation period. 10.2 Test Instruments Description & Manufacturer TESEQ 5KVA AC POWER SOURCE/CONDITION UNIT Software Model No. Serial No. NSG 1007/ CCN 1000-1 1323A00070/ 1318A02119 Win2100 V4.5.8 NA Cal. Date Cal. Due Aug. 21, 2014 Aug. 20, 2015 NA NA Notes: 1. The test was performed in Hwa Ya EMS Room. 2. The calibration interval of the above test instruments is 12 months and the calibrations are traceable to NML/ROC and NIST/USA. 10.3 Test Arrangement a. The EUT was placed on the top of a wooden table 0.8 meters above the ground and operated to produce the most unfavorable sequence of voltage changes under normal operating conditions. b. During the flick measurement, the measure time shall include that part of whole operation cycle in which the EUT produce the most unfavorable sequence of voltage changes. The observation period for short-term flicker indicator is 10 minutes and the observation period for long-term flicker indicator is 2 hours. EUT Report No.: CE150723C22 Page No. 48 / 87 Report Format Version: 6.1.2