Download Manual - Astronics Test Systems

Transcript
Publication No. 980673-043 Rev. A
1260-39 User Manual
SLIST
Relay descriptors are always unique for each 1260 module. The
following paragraph describes the relay descriptor syntax for the
1260-39 module.
1260-39 Relay
Descriptors
A "relay descriptor" identifies a relay (or range of relays) to be
operated. The "relay descriptor" is unique for each 1260 module
type.
The "relay descriptor" for the 1260-39 has the form:
<module address>. <channel range>
Where:
<module address>
inclusive.
is an integer between 1 and 12,
NOTE:
The <module address> used here is not the VXIbus
defined Logical Address of the 1260 Series Master. It
is unique to the 1260 Series and describes the
switching module in relation to the Master. This
address corresponds to the binary value of the switch
setting of SW1 on the switching module PCB. Refer
to the Installation Section of the 1260 Series VXI
Switching Cards Manual for more information.
<channel range>
is a single relay, or a list of relays. A
comma or a hyphen may be used to separate relays in a
<channel range>. When a comma is used, only the specific
relays are operated; when a hyphen is used, all relays
between the relays are operated.
Each relay is identified by a single 4-digit channel number using
the format:
<1 digit relay type> <1 digit selector> <2 digit relay
identifier>
Astronics Test Systems
Module Specific Syntax 3-3