Download User Manual COMBINATION 855

Transcript
TENS
Waveform
Pulse Width
Repetition Rate
Output type
Diadynamic Currents
Current Types
Output type
Sinusoidal
Frequency (AMF)
Surge Rate
Surge Pattern
Output Type
Faradic
Frequency
Surge Rate
Surge Pattern
Output Type
Galvanic
Output Type
Interrupted Galvanic
Pulse Width
Waveform
Pulse Rate
Output Type
Träbert
Waveform
Output Type
Asymmetrical, Symmetrical or Sequential
20 – 400 µs in 5 µs increments (±10%)
1 – 250 Hz in 1 Hz increments (±10%)
CC 0-100mA peak (±10%) into 500 ohms
CV 0-60V peak (±10%) into 2 kohms
DF, MF, CP, CPiso, RS, LP
CC 0-70mA peak (±10%) into 500 ohms
CV 0-140V peak (±10%) into 2 kohms
50 Hz (±5%)
2 to 30 /minute (±10%)
Rectangular, Triangular or Trapezoidal
CC 0-70mA peak (±10%) into 500 ohms
CV 0-140V peak (±10%) into 2 kohms
50 Hz (±5%)
2 to 30 /minute (±10%)
Rectangular, Triangular or Trapezoidal
CC 0-70mA peak (±10%) into 500 ohms
CV 0-140V peak (±10%) into 2 kohms
CC 0-70mA peak (±10%) into 500 ohms
CV 0-140V peak (±10%) into 2 kohms
10 µs to 1 s (± 5%) for rectangular in 1, 2, 5
sequence
1 ms to 1 s (± 5%) for other shapes
Rectangular, Triangular or Trapezoidal
1 to 30 /minute (±10%)
CC 0-70mA peak (10%) into 500 ohms
CV 0-140V peak (±10%) into 2 kohms
2 ms on (± 5%), 5ms off (± 5%) rectangular
CC 0-70mA peak (±10%) into 500 ohms
CV 0-140V peak (±10%) into 2 kohms
Combination 855