AMETEK Programmable Power
BPS / MX / RS Series SCPI Programming Manual
DO160:NORMal:VOLTage:MODulation
This command will cause output voltage modulation. The level of modulation is the function
of the rate of modulation. Refer to Figure 9-4 in the User Manual. The command parameter
is the rate of modulation in Hz.
Command Syntax
Parameters
Examples
Query Syntax
Returned Parameters
DO160:NORMal:VOLTage:MODulation <NRf>
<frequency>
DO160:NORM:VOLT:MOD 20
DO160:NORM:VOLT:MOD?
<NR1>
DO160:NORMal:VOLTage:INTerrupt
This command will cause output voltage interruption. There are 15 different levels of
interruption. Refer to Figure 9-6 in the User Manual for characteristic of each level.
Command Syntax
Parameters
Examples
Query Syntax
Returned Parameters
DO160:NORMal:VOLTage:INTerrupt <NRf>
<test number>
DO160:NORM:VOLT:INT 4
DO160:NORM:VOLT:INT?
<NR1>
DO160:NORMal:VOLTage:SURG
This command will generate the voltage levels required to generate a normal voltage
source. Refer to Section 9.1.4.1, under the heading titled VOLTAGE SURGE, for detail.
Command Syntax
Parameters
Examples
Query Syntax
DO160:NORMal:VOLTage:SURG
none
DO160:NORM:VOLT:SURG
none
DO160:NORMal:VOLTage:DC:UNDer
This command will generate the voltage levels required to generate a normal voltage
source. Refer to Section 9.1 in the User Manual under the heading titled VOLTAGE UNDER,
for detail.
Command Syntax
Parameters
Examples
192
DO160:NORMal:VOLTage:DC:UNDer
none
DO160:NORM:VOLT:DC:UND
Manual P/N 7003-961 Rev AA