conditions, and run ACQUITY UPLC system diagnostic tests. These basic
functions are rapidly performed, even when a system is remote from the
software control and acquisition workstation or LAC/E™32 module or when
network control is unavailable.
FlexCart
The optional FlexCart provides for the ACQUITY UPLC H-Class system a
mobile platform. It can hold the system instruments as well as the PC and
monitor and provides electrical outlets for system instruments and integrated
waste management. Used with a mass spectrometer, the cart’s adjustable
height lets you position the column outlet close to the inlet probe, minimizing
system dead volume.
Column technology
ACQUITY UPLC columns are packed with 1.7-μm, bridged, ethylsiloxane,
hybrid or 1.8-μm high strength silica particles that can mechanically endure
high-pressure conditions. The column hardware and the matched outlet
tubing can withstand up to 103,421 kPa (1034 bar, 15,000 psi). The column
dimensions allow optimal MS-compatible flow rates, and matched outlet
tubing minimizes the effect of extra-column volume.
Although the system works with any analytical HPLC column, specially
designed ACQUITY UPLC columns maximize its high-pressure capabilities.
Compared with traditional HPLC columns, ACQUITY UPLC columns deliver
superior resolution and sensitivity in the same run time, or equivalent
resolution, greater sensitivity, and faster run times.
eCord technology
ACQUITY UPLC columns include an eCord column chip that tracks the usage
history of the column. The eCord column chip interacts with the system
software, recording information for up to 50 sample queues run on the column.
In regulated environments, the eCord column chip provides documentation of
the column used in the validation method.
System components