Download ACQUITY UPLC H-Class System Guide

Transcript
Connecting to the electricity source
Each system instrument requires a separate, grounded power source. The
ground connection in all power outlets must be common and physically close to
the system.
Warning: Avoid electrical shock:
• Use power cord SVT-type in the United States and HAR-type or
better in Europe. For other countries’ requirements, contact your
local Waters distributor.
• Power-off and unplug each system instrument before performing any
maintenance operation on the instrument.
• Connect each system instrument to a common ground.
To connect to the electricity source
Recommendation: Use a line conditioner and uninterruptible power supply
(UPS) for optimum, long-term, input voltage stability.
1.
Connect the female end of the power cord to the receptacle on the rear
panel of each instrument.
2.
Connect the male end of the power cord to a suitable wall outlet.
Alternative: If your system includes the optional FlexCart, connect the
female end of the Flexcart's electrical cables (included in the startup kit)
to the receptacle on the rear panel of the each instrument. Connect the
hooded, male end of the Flexcart's electrical cables to the power strips on
the back of the cart. Finally, connect each power strip's cable to a wall
outlet operating on its own circuit.
Connecting to the electricity source