Download Yearly Maintenance
Transcript
0123004 !"#$ 56789 01232456 01232456,0178 %&'()*+,-./ Table of Contents Note: Pages 3 - 4, 10 - 12 and 16 - 17 are omitted in this document 1 General………………………………………………………………………………….………………………………..5 2 Visual Check…………………………………………………………………………………..………………..……..6 3 Indicator Lights and audible Signal……….…………………………………………………….….……….7 4 Power supply adequacy…………………………………………………………………….………….…………9 5 Yearly Calibration Documentation: Please Fax Copy to Sirona 1-888-297-8631……...13,14,15 6 Adjusting / calibrating the system………………………………………………….…………….……....18 7 9.1 System adjustment menu…………………………………………………...18 9.2 Diaphragm Image/GALILEOS diaphragm adjustment….….…...20 9.3 Diaphragm image/GALILEOS GAX5 diaphragm adjustment.…23 9.4 Checking the beam field……………………………………………….....….26 9.5 Dosimetry……………………………………………………………………..…….27 9.6 Sensor calibration……………………………………………..…………………28 9.7 Iris calibration……………………………………………...…….…………...….29 9.8 Shading calibration……………………………………….……………..…..….30 9.9 Distortion calibration…………………………………………..………….…..31 9.10 Geometric calibration…………………………………….……………...……32 Checking the laser……………….……………………..…………………………………………..…….………33 012345678495 60986 514984548649291346 0 WC×A?AF=P bX 8923855415 63 8634982c6378 6d00e6 06d26)($6^"63 8c6326861L5 849'' ¦H§AFC̈G©©PªC«RAQG=Bª 2365286955969586 06986382832548616146 532754864195601234563818159134c654657865986334826 33 6832638254849 61495 59134616884915 ' 2586I87825 6J8283254860954752768326715943916h:256 419c63786386I87825 6e89 59134c6219 86)"6[Iec60553: 9826¬c654759865465786598633482633 ' G@ªC®ªBPA ¯89488468°3328651495146596 85965633 :38869186i593: 59168°33286 3Y598c68863825914964929134654: 5 j' ±©AF=@>?²CD?E@FGB@>H?E 72149685288496547634923 6196164885269368482: 91³8632678:848291³869586419'6I3265 6338259149698363 8586 288826936958638259149649291346545 ' ®«´ND±µC§>@OC¦®C¶H=FRE· 6J[63527652868199876419568 892341633348496841: 91L869368 89239591671552986i07j'64654684L12348496386 3L14963833 868 892395916552986528645L3175 86786936 8219134658766 395149c652389149689' +,-./01 23695614633 1548641956958670628612884969586 06968651495148765445 683 34149675986386 1495 59134' 961695862833411 1963869586826936142869559695868613: 849616514951487641956958654859282;6283847876 <=>?@A?=?BACD?E@FGB@>H?E69361428633 15486419569586 I87825 6J82832548609547527' 25865485928265476958658 82K1495 826528628 18L876 82362833411 1961469538658645828643433 15486 4195695869547527628 968236958682;6851 28693655L869586 54859282;6283847876514984548638283287' 2586595 6514984548614389134654763486849682: L1869686533 158766569251487682L1854'6M81: 958269586143891346432682L1861635296386958686138496452: 2549' NABO?>B=PC>?E@FGB@>H?ECFAQG>FAR 3825914964929134 082L186545 D?E@FGSA?@EC=?RC=RTGE@SA?@C@HHPECFAQG>FAR 6852149678L1869686195 86832685288496 3469586514633984915 ' 251614 78695863 8634982' 2516854U66L3 95986386"(((6V68948846563448913465476 9234769686331 8' WX 719195 6 9189826I Y86#&6ZZZ6326861L5 849' 25U 7[6L3 959\]^6('"6_6386285714963 6('()_638625498 7[6228`a]^6('%6_6386285714963 6('"_638625498' ¸l¹srtm {yxzx|{~~ x{º{xx|y{}w{{|{|}{w~x »¼~w{}~x{y|{v~yw{v{|}xy{x xw ½lpmrmn x|y~{¾}¿{À~Á~y »¼~w{|y{|{uÃÃ{vxºyx{x| {~{wx| { |}x{|xw|{x~w{|{|}x{|xw|{|w M II fX 7g98496896419565 1948496933 68326h:256856c69896 3Y654764887 86355493678 1L82876419569586419c6i: klmnop 9382;63233829j' ÄÅÆÇÈÉÊËÇÌÉÇÍÅÍÎqÏÐÑÉÌÎÇqÐÍÒÓÔÕÖ klmnopqplkrlsrtm uvwxyzx{|}x{y~~|{y|x|{ x{xw{~w{|x{{ |}x{uxy~| {w|y|w{~~ y~~|{w{x||x{~w{ {~w{|}x{yxx~wx{v||{w{x yxwwx }x{y~{~|y{w|{ }|{{{|}x{~w~{ ¡ ¡¢¡¡£¤¥¤ !"6"!6#"$67$%$& 7$%$&'"($'("'($'() * 0123456789 77 7 UKVIEWGHKXYMZ[ !"##$% $\ 2> #5 0 #12345622547 2- 8 92322:44524!64%42;44<= >55? 8 @!A 8 @!"44 "4A!>3A!53 !2426!6-B! CDEFGHHGFIJDKLMNJOFPQGOOGDFRKSGEENJOFT , &''&('))*)+ )*)+-'.)-.'-.)-./ 012345678495 60986 %51498454&86'492(&91346)*'*+,0 mCN@cA]\BdZCFA_`BXC\@cC\WcAnoYCpA_@\o !"#$" s8 8586(9934 s5Q159134614Q1&5932 U4196,O6*+7 ,O ,tt u\c vwxyz s8398 k23V3 ^ 789:;<= G\Xbu\c ir ) q > ?@ABCDECFGHI 78J2869K865146L19&K6149369K86MNM6J31913469369(246(4196 ,O36 PK86(41965QR(96198 S65(93591&5 T6L519653(96 .14(983 PK86MU4196,O6*+7M61469K86(JJ82621VK96&3248263S69K86WXYZC A@BYZ[\]YCL1 69K84614Q1&59869K5969K86(419616,O3 > ^YXBA@_CB`YCaYbXC\@cCB`YC[W@]BAd@XCd@CB`YCWXYZCA@e BYZ[\]Y3 f gK8469K86K81VK965QR(98496165Q869K86 3(Qh J85i826K3( Q6K(3 f 'S6283986161495 8QT698969K86283986i8654Q6Q1h J 561465QQ191343 f PK8628 8586(99346(96439686Q8S8&91j8632623i843 08865 36,J825914V6'492(&91346(4Q826Mk28J5214V69K86+lJ3h (28M6(&K5J9826M0L19&K14V69K8698634M3 -.6.-6/.0670102 701023.4034.340345 6 0123456789 47 623 634 426 ! "!#$ %&'(&")*'*+, ³¶À¼»½¾°Á»º¶°Ã»¾¶µ ®¯° ª¬ ±ª² ®¯° ©ª«¬ ½¼ ¿· ½¾° » ¼ ¹ º» ±¹ ° ¶ ¸ · ¶ ®´µ W>;B9ÄÅ<@BAÆÄ?@ NHH& §,¨*+ ¦U4 ¤ N(MN Y<ZM (#!( ÇÈÉÉÊWË §,¨*+ NHH& M\ !) ¥ N(MN Y<ZM (#!( ÇÈÉÉÊWË°ÁªÌÍ 7 89:;<9<;=>?@AB;C D EFGH!(#GIJ'KL I (MNO4 7 ;BBA><9<;=>?@AB;<Y<;9 9<;9AB; D + #GPI(H Q'+E' RS4 D ,#Q&FKK!K!P4!4# T!HHI #! U PI(J(!&PI(#&MUO4 ,IU'(&4 D U!#Q4 D G<VWXPI(IU!( UGY<ZM4 D &3[M\]X^!) (UY<ZM4 _`abcded`_f̀gfha i _jkljmmenojeljwjumjesrnnqtertnxwelu{ewxonmerkeutve mrsmjrjtnw{eljwjumjeienojejpkqmrljeyrmne njlyxtunjexyyjvxunjw{|egojeluvxunxqtextvxzunqle _`abcded`_f̀gfha wxonmerk| i _jkljmmenojejpkqmrljesrnnqteutveoqwvertnxwenoj jpkqmrljenjlyxtunjmeurnqyunxzuww{|e i gojejpkqmrljejtvme}ojtenojelqnunxqteutveluvxunxqte _`abcded`_f̀gfha urnqyunxzuww{em}xnzoeq~~| _j ~ j z wxonextvxzunqlm i gojeluvxunxqtextvxzunxqteyrmnewxonerkevrlxtenoje nxje ¡ jpkqmrljekjlxqv|e goj e rm j l e x m e tqn i xyrwnutjqrmw{euteurvxswjesjjkeyrmnemqrtveunenoje uljeyuvj£ ekjlyxnnjvenqermjenojertxn¢ertnxweljkuxlme rtxn| 3 -..-/.00102 01024.5045.450456 012345678495 60986 !51498454"86#492$"91346%&#&'(0 lFÄHÅENFqsOOBÆFKMEÇsK Æ RS \S bÉÉFÊËÌ bS d Èj Èj g23P/ % Í _R h] % i : ;36<89821486=3>826$== 65<8?$5"@69A86BCDEFGHBIJ KLEFMNHO6<$214P68Q=3$286$968685$28</ RS T86$286=3>82616<1"3448"98<65969A86"84925 6<1921$U 91346=548 V6 : W83X86Y23496"3X826ZY326<8951 6886082X1"86!54$5 [/6 \S 08 8"96,00]%^6 1486X3 95P86254P86346$ 9189826/ ^3448"9685$214P6 85<6936982145 6_R@6̀654<6a/ bS ^3448"96=3>82654<6>19"A6$4196(c6d/ e5196*614/6Y3268 YU5<f$984963Y69A86$419/ g286h86iFj6936289$246kU2569$86A85<6149369A86141915 6 =319134/ lS '95 1A68Q=3$286285<1486X1560#7'k#0mn/6 oS 08 8"96pCLpEqIFErOHqsNEFBEGEBF8/P/6tuh]vwt% xyz{|}~ )*6*)6+*,67,-,. 7,-,./*0,/0*/0,/01 3456789 d _R &1486X3 95P8@6436 35<Î !5Q/6=8211 86 1486X3 95P86 <23=Î Ð] *+06Ï610+] +] 10+6Ï61,0] ./Ñ] 1,06Ï61-0] .] 1-06Ï61)-] iE HNMFNEKMCDL/6 : ;$246$4196(¡¡/66 : W83X8689826 85<654<628Y196"3X82/ ¢ £¤¥¦£§¨©ª «¬®§°̄®§±²³̄´®µ¶§°̄¨¯§¨·§¨©®¸²¨¯®§¹³º®µ§²¹¹»¼§´²¯§½®§ ¬·¯¨»»®©¾§¿®À®µ§³̄§Áµ®ªÂ·¯¨»»¨¯¬³·§Â·µ̄²±¯¬³·¾§ ¦¦£Ã 2 03=A9 1=323354B56718469:54;9<=1:71;>31?35@9831 012345678495 60986 852 65 12591346698682186841488263269860628826936986514984548649291346 CDEFGDHIJKLLKMNHOIPIQROESIFKITNDKHRU V3=354619=W<5X4;9<=1W<51;>31?35@98313=A9=335 Y86ZZ[\06]35@9831^4=_466\2`826a3%6 6!6!61614`1b845 86326b82321469865 1259134%6 636986 8528628c128`63269865 125913465286̀8218`698286146986d5b982ed5 1259146986419%f g2982328h6986\b825914649291346i165868846bb 18`69389826i1969864196528628c128`% n )*+,-./ jklm o=14pp9;9<=1;<1;>314==_4618>38q1:71;>31?7?;3X1<r=351<51<;>3514_;><59s3p1t35?<=?u14173456718469:54;9<=1 X_?;146?<1:31t35W<5X3p1:71;>31?35@98313=A9=335v 03 `61968648852693628b 586̀8891863b3484965 346i196b5296w896936i852h691696865288`6 b2813 6i19698693826326̀8495 6b25918695% 6 xyz{|z}~z~}x{z|}}|~{} {} 713491459134636b5286b5296̀8 18218 g32685698h6986b5286b5296̀8 1821865286̀1349148`659826568295146b8213`636918%6Y869863i4826i1 6 8628b341 8632658928 8549651 28636986i1634914861463b82591346598269596918654`6546436 34826868218`6̀869361146b5286b529% _?;<X351p4;4 _?;<X35 4?;1=4X3 95?;1=4X3 ];533; <?;18<p31189;7 ><=3 0b586326b25918695b 6!6!""676"#"$ 76"#"$%&!%& %&"%& '( 01234567248932 86956 1613871618113636 16631136 6 1628 12716 37 86 83261 2465 169 &'()*+',)-./,01/23+'4/5-), 677177)-./23+'4/189:75+17 @',1/1 136 A5123 1 238782741662 8 62B2C46 1 6DEF6 11761 8 3236 31264DEF 121G61 8711H6G2 6 6 I1 11266 183148B761DEF 626246 11861 13 35876DEF 1626246 1367248932 86DEF6 1161387162246 3867248932 86DEF6 1161387162246 JKLMNOPQ7248932 86DEF 1161387162246 RNSTUVSNUOQ7248932 86DEFQ 1161387162246 11 38767248932 8QWXFQ 1161387162246 Y248932 866 773116DEFQ 1161387162246 6-7;1+/<517,):-7/;),0/417/=>/:+/-:/=?> 66Z6Y3317 624 [ % 14 6!6!""66"#"$ 012345678495 60986852 65 12591346698682186841488263269860628826936986514984548649291346 ()*+,-./0.12/3-24-56742*570865/0). 9864:82148:63412695968;86568<8:6986419632698653862198215654:695968;865654:8:6 38269864196146 6491345 634:19134== HF IF JF KF LF MF NF OF PF HQF HHF HIF HJF HKF B0 ),/60*2 C35 ./), D-60572E)FA (0 G/./ )6), G0).2 E5,-A C35 D-60572E)FA (-57-6A D01.5/+6-A !"#$%&' >-56 (5 *3-/-*2@)? A 15 9 Adjusting / calibrating the system 9.1 System adjustment menu 9 Sirona Dental Systems GmbH Maintenance Instructions GALILEOS Adjusting / calibrating the system 1. 2. 3. SERVICE menu tab Message window Preview Image Tools pictogram 9.1 System adjustment menu The SERVICE FUNCTIONS menu (1.) will guide you through the calibration process. This service routine is started from SIDEXIS XG: UTILITIES Æ CONSTANCY TESTÆ 3D Æ (SELECT X-RAY DEVICE) Æ SERVICE EXPOSURE Æ password query Æ (SELECT X-RAY COMPONENT) Æ SERVICE FUNCTIONS menu The queries SELECT X-RAY DEVICE and SELECT X-RAY COMPONENT will only appear if several units are set up in SIDEXIS XG. i NOTE The SERVICE FUNCTIONS menu is password-protected. As password, enter the first four digits of the current system date (PC) in reverse order. Example: On 05/30/2004, the service password is 5003 i NOTE When you open the SERVICE FUNCTIONS menu, the unit switches from the user mode to the PC service mode logged by the PC. For GALILEOS: This mode is displayed on the Easypad via the PC service image (2.). For GALILEOS GAX5: This mode is displayed on the Multipad via the "SERVICE" display (3.). . 18 Status column In PC service mode, the control options that are available on the control panel are determined by SIDEXIS XG and the currently selected service routine. General control of the unit by means of the control panel (as in the user mode) is not possible in this mode. The SERVICE FUNCTIONS menu contains 9 sub-menus that can be selected using the menu tabs. The tools pictogram shows which (if any) test phantom should be used for this particular adjustment step. The message window displays text messages regarding the adjustment process.. To the right of the menu you can see the status column. This column contains information about the system's current calibration state. 9 green and checked valid data record, calibration is in progress. green calibration data record present, calibration has not been performed, calibration might not be sufficient. No calibration required! Calibration required! red invalid data record or no record present Calibration required! To quit the SERVICE FUNCTIONS menu, click CANCEL. 61 16 813 D3437 D3437.103.01.04.02 Sirona Dental Systems GmbH Maintenance Instructions GALILEOS GALILEOS 9 Adjusting / calibrating the system 9.1 System adjustment menu GALILEOS GAX5 Preview image båÖäáëÜ The DIAPHRAGM and SHADING submenus each contain a preview image that symbolizes the exposure to be taken during the calibration step. Due to the varying geometry of the GALILEOS and the GALILEOS GAX5 diaphragms, the preview images displayed in these sub-menus differ slightly (see illustrations above). We use only the display of the GALILEOS in these instructions, unless explicit reference is made to the GALILEOS GAX5. To quit the SERVICE FUNCTIONS menu, click CANCEL. 61 16 813 D3437 D3437.103.01.04.02 19 9 Adjusting / calibrating the system 9.2 Diaphragm exposure Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 3. R 1. WAIT! UNTIL UNIT IS IN STARTING POSITION 4. 2. 5. C A Adjustment: not OK B Adjustment: OK 9.2 Diaphragm exposure 5. 9.2.1 GALILEOS diaphragm adjustment „Type 2“ • Open the SERVICE FUNCTIONS menu. 1. Select the DIAPHRAGM sub-menu. 2. Make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S030.5 is displayed on the Easypad touchscreen. 3. Press the R key to move the unit back to the starting position. 4. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal for end of exposure (double beep) sounds. 20 i Evaluate the image. – The brightness distribution along the border surrounding the image on all sides must be uniform (A). – The distance from the lower edge to the lowest point in the image must be within the tolerance range (inside the auxiliary lines) (B) NOTE If the distance between the bottom edge and the lowest point in the image is out of tolerance (B) or the brightness distribution along the surrounding border is not uniform (C), the diaphragm must be adjusted mechanically (see Service Manual). 6. If the exposure is OK (A+B), confirm this by clicking the check box underneath the message window. The box will appear checked. 61 16 813 D3437 D3437.103.01.04.02 9 Adjusting / calibrating the system 9.2 Diaphragm exposure Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 3. R 1. WAIT! UNTIL UNIT IS IN STARTING POSITION 4. 2. D C A Adjustment: OK Adjustment: not OK: diaphragm not centered båÖäáëÜ 5. Adjustment: not OK: diaphragm too small B 9.2.2 GALILEOS GAX5 diaphragm adjustment „Type 2“ • Open the SERVICE FUNCTIONS menu. 1. Select the DIAPHRAGM sub-menu. 2. To make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION – The distance from the lower edge to the lowest point in the image must be within the tolerance range (B) – The brightness distribution along the surrounding border must be uniform (A). A shadow in the image indicates horizontal or vertical displacement of the diaphragm. – No surrounding gray shadow should be visible (D). A surrounding gray shadow in the image indicates that the diaphragm is too small. The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S030.5 is displayed on the Multipad. 3. Press the R key to move the unit back to the starting position. 4. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal that indicates the end of the exposure (double beep) can be heard. 5. Evaluate the image. – The brightness distribution along the border surrounding the image on all sides must be uniform (A). 61 16 813 D3437 D3437.103.01.04.02 i NOTE If the distance between the bottom edge and the lowest point in the image is out of tolerance (B) or the brightness distribution along the surrounding border is not uniform (C) or a surrounding gray shadow is visible in the image (diaphragm opening too small) (D), the diaphragm must be adjusted mechanically (see Service Manual). 6. If the exposure is OK (A+B), confirm this by clicking the check box underneath the message window. The box will appear checked. 21 9 Adjusting / calibrating the system 9.2 Diaphragm exposure Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 2. 4. R 1. WAIT! UNTIL UNIT IS IN STARTING POSITION 5. 3. 6. A C Adjustment: not OK B Adjustment: OK 9.2.3 GALILEOS diaphragm adjustment „Type 3“ • 1. 2. 3. 4. Open the SERVICE FUNCTIONS menu. Select the DIAPHRAGM sub-menu. The DIAPHRAGM OPEN button in the IMAGE ACQUISITION group is selectable. "Open diaphragm" diaphragm setting Set the rotary knob on the tube assembly to the "open diaphragm" position. Bring SIDEXIS XG into exposure readiness for the exposure in the "open diaphragm" diaphragm position: Click the DIAPHRAGM OPEN button. The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S030.5 is displayed on the Easypad touchscreen. Press the R key to move the unit back to the starting position. 5. 6. i NOTE If the distance between the bottom edge and the lowest point in the image is out of tolerance (B) or the brightness distribution along the surrounding border is not uniform (C), the diaphragm must be adjusted mechanically (see Service Manual). 7. 22 Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal that indicates the end of the exposure (double beep) can be heard. Evaluate the image. – The brightness distribution along the border surrounding the image on all sides must be uniform (A). – The distance from the lower edge to the lowest point in the image must be within the tolerance range (inside the auxiliary lines) (B). If the exposure is OK (A+B), confirm this by clicking the check box underneath the message window. The box will appear checked. 61 16 813 D3437 D3437.103.01.04.02 9 Adjusting / calibrating the system 9.2 Diaphragm exposure Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 2. 1. 4. R WAIT! UNTIL UNIT IS IN STARTING POSITION 5. 3. D C A Adjustment: OK Adjustment: not OK: diaphragm not centered båÖäáëÜ 6. Adjustment: not OK: diaphragm too small B 9.2.4 GALILEOS GAX5 diaphragm adjustment „Type 3“ • 1. 2. 3. 4. 5. Open the SERVICE FUNCTIONS menu. Select the DIAPHRAGM sub-menu. The DIAPHRAGM OPEN button in the IMAGE ACQUISITION group is selectable. "Open diaphragm" diaphragm setting Set the rotary knob on the tube assembly to the "open diaphragm" position. Bring SIDEXIS XG into exposure readiness for the exposure in the "open diaphragm" diaphragm position: Click the DIAPHRAGM OPEN button. The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S030.5 is displayed on the Easypad touchscreen. Press the R key to move the unit back to the starting position. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal that indicates the end of the exposure (double beep) can be heard. 61 16 813 D3437 D3437.103.01.04.02 6. i Evaluate the image. – The brightness distribution along the border surrounding the image on all sides must be uniform (A). – The distance from the lower edge to the lowest point in the image must be within the tolerance range (B) – The brightness distribution along the surrounding border must be uniform (A). A shadow in the image indicates horizontal or vertical displacement of the diaphragm. – No surrounding gray shadow should be visible (D). A surrounding gray shadow in the image indicates that the diaphragm is too small. NOTE If the distance between the bottom edge and the lowest point in the image is out of tolerance (B) or the brightness distribution along the surrounding border is not uniform (C) or a surrounding gray shadow is visible in the image (diaphragm opening too small) (D), the diaphragm must be adjusted mechanically (see Service Manual). 7. If the exposure is OK (A+B), confirm this by clicking the check box underneath the message window. The box will appear checked. 23 9 Adjusting / calibrating the system 9.2 Diaphragm exposure Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 8. 10. R WAIT! UNTIL UNIT IS IN STARTING POSITION 11. 9. 12. A B B Adjustment: not OK A Adjustment: OK 9.2.5 Diaphragm adjustment „Type 3 Upper jaw" 1. Set the rotary knob on the tube assembly to the "maxillary exposure" position. 2. Bring SIDEXIS XG into exposure readiness for the exposure in the "upper jaw" diaphragm position: Click the UPPER JAW button. The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S030.5 is displayed on the Easypad touchscreen. 3. Press the R key to move the unit back to the starting position. 4. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal that indicates the end of the exposure (double beep) can be heard. 24 5. i Evaluate the image. – The distance from the upper edge of the lower lead diaphragm must be within the tolerance range, i.e. inside the auxiliary lines (A). NOTE If the edge is outside of the tolerance range (B), the diaphragm must be adjusted mechanically (see Service Manual). 6. If the exposure is OK (A), confirm this by clicking the check box located to the right of the UPPER JAW button. 61 16 813 D3437 D3437.103.01.04.02 9 Adjusting / calibrating the system 9.2 Diaphragm exposure Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 14. 16. R WAIT! UNTIL UNIT IS IN STARTING POSITION 17. 15. A C B båÖäáëÜ 18. B Adjustment: not OK A Adjustment: OK 9.2.6 Diaphragm adjustment „Type 3 Lower jaw" 1. Set the rotary knob on the tube assembly to the "mandibular exposure" position. 2. Bring SIDEXIS XG into exposure readiness for the exposure in the "lower jaw" diaphragm position: Click the LOWER JAW button. The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S030.5 is displayed on the Easypad touchscreen. 3. Press the R key to move the unit back to the starting position. 4. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal that indicates the end of the exposure (double beep) can be heard. 61 16 813 D3437 D3437.103.01.04.02 5. i Evaluate the image. – The distance from the lower edge of the upper lead diaphragm must be within the tolerance range, i.e. inside the auxiliary lines (A). NOTE If the edge is outside of the tolerance range (B), the diaphragm must be adjusted mechanically (see Service Manual). 6. If the exposure is OK (A), confirm this by clicking the check box located to the right of the LOWER JAW button The box will appear checked. 25 9 Adjusting / calibrating the system 9.3 Checking the radiation field 1. Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 2. 2. 3. A 4. B R WAIT! UNTIL UNIT IS IN STARTING POSITION 5. 9.3 Checking the radiation field i NOTE Illumination must be checked once the diaphragm has been adjusted. 1. Clip distortion phantom A onto the X-ray detector cover. 2. Go to the RADIATION FIELD sub-menu. 3. Make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S002.6 is displayed on the control panel. 4. 5. 26 Press the R key to move the unit back to the starting position. Press the release button. Hold down the release button while checking the distortion phantom. Lighting strips B on the distortion phantom must not light up! i NOTE If the strips on the phantoms light up at all it means overexposure of the system, and you cannot continue adjustment. In this case, repeat the diaphragm adjustment procedure and then check the radiation field again. If the lighting strips still light up during the re-check of the radiation field, contact the SIRONA Customer Service Center (KSC) to solve the problem. To confirm that the lighting strips on the distortion phantom do not light up, click the check box underneath the message window. • The box will appear checked. • Continue the calibration procedure with dosimetry. i NOTE Leave the distortion phantom on the unit for the next calibration step. 61 16 813 D3437 D3437.103.01.04.02 9 Adjusting / calibrating the system 9.4 Dosimetry Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 2. 1. 4. båÖäáëÜ 3. R WAIT! UNTIL UNIT IS IN STARTING POSITION 5. 9.4 Dosimetry i 6. 5. NOTE A dose measurement device (e.g. Mult-O-Meter type 510L) is required for dosimetry. The distortion phantom should be clipped onto the X-ray detector cover during the dose measurement process for protection against scratching. 1. Go to the DOSE sub-menu. 2. Attach the Mult-O-Meter sensor approximately in the middle of the distortion phantom mounted on the X-ray detector. 3. Make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S002.6 is displayed on the control panel. 4. i Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal for end of exposure (double beep) sounds. Then read the dose on the Mult-O-Meter. The value must be between 1.2 and 2.3 mGray. NOTE If the value is outside the permissible range (1.2 to 2.3 mGray), check the tube (see Service Manual). 6. • • • To confirm that the dose is within the permissible range between 1.2 and 2.3 mGray, click the check box underneath the message window. The box will appear checked. Remove the sensor from the distortion phantom again and remove the phantom from the X-ray detector. Dosimetry is now completed. Continue the calibration procedure with sensor calibration. Press the R key to move the unit back to the starting position. 61 16 813 D3437 D3437.103.01.04.02 27 9 Adjusting / calibrating the system 9.5 Sensor calibration Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 3. R 1. WAIT! UNTIL UNIT IS IN STARTING POSITION 4. 2 - 3 minutes! 2. 5. 6. 9.5 Sensor calibration 1. Go to the SENSOR sub-menu. 2. Make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S010.14 is displayed on the control panel. 3. i If you have repeated the procedure three times and still have not attained a positive result, please contact the SIRONA Customer Service Center. Press the R key to move the unit back to the starting position. 6. 4. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal for end of exposure (double beep) sounds − this takes approx. 2 − 3 minutes. 7. • 5. The acquired images are transferred from the unit to the Reconstruction and Control Unit (RCU). This may take 2 − 3 minutes. The evaluation of the sensor calibration is then displayed in the message window. 28 NOTE If the information in the message window indicates that calibration is not ok and/or not possible, keep repeating the procedure starting with Step 2 until calibration is ok and/or possible. If calibration is ok and/or possible (see above), click SAVE VALUES to save the calibration. Sensor calibration is now completed. Continue the calibration procedure with iris calibration. 61 16 813 D3437 D3437.103.01.04.02 9 Adjusting / calibrating the system 9.6 Iris calibration Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 3. R 1. WAIT! UNTIL UNIT IS IN STARTING POSITION 4. 2 - 3 minutes! 2. båÖäáëÜ 5. 6. 9.6 Iris calibration 1. Go to the IRIS sub-menu. 2. Make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S010.10 is displayed on the control panel. 3. Press the R key to move the unit back to the starting position. 4. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal for end of exposure (double beep) sounds this takes approx. 2 − 3 minutes. 5. The acquired images are transferred from the unit to the Reconstruction and Control Unit (RCU). This may take 2 − 3 minutes. The evaluation of the iris calibration is then displayed in the message window. 61 16 813 D3437 D3437.103.01.04.02 i NOTE If the information in the message window indicates that calibration was not successful, keep repeating the procedure starting with step 2 until calibration is ok. If you have repeated the procedure three times and still have not attained a positive result, please contact the SIRONA Customer Service Center. 6. • • If calibration was successful, click SAVE VALUES to save the calibration. Iris calibration is now completed. Continue the calibration procedure with shading calibration. 29 9 Adjusting / calibrating the system 9.7 Shading calibration Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 3. R 1. WAIT! UNTIL UNIT IS IN STARTING POSITION 4. 2. 5. 6. 9.7 Shading calibration 1. Go to the SHADING sub-menu. 2. To make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in SIDEXIS XG. Service routine S010.11 or S010.15 (extended shading calibration) is displayed on the control panel. i NOTE If the message window indicates that the calibration was not successful, repeat the procedure starting with step 2 until the calibration is OK. If you have repeated the procedure three times and still have not attained a positive result, please contact the SIRONA Customer Service Center. 3. Press the R key to move the unit back to the starting position. 4. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal that indicates the end of the exposure (double beep) can be heard. No foreign bodies may be visible on the shading exposure. If this is the case, check the beam path for foreign bodies and remove them if applicable. Repeat the calibration. 5. The Shading exposure and the evaluation of the Shading calibration are displayed in the message window. 6. CAUTION • • 30 If calibration was successful, click SAVE VALUES to save the calibration. Shading calibration is now completed. Continue the calibration procedure with distortion calibration. 61 16 813 D3437 D3437.103.01.04.02 9 Adjusting / calibrating the system 9.8 Distortion calibration Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 3. 1. 2. 4. 5. båÖäáëÜ A R WAIT! UNTIL UNIT IS IN STARTING POSITION 6. 9.8 Distortion calibration 1. Go to the DISTORTION sub-menu. 2. Read off the serial number of distortion phantom A from the ID label of the phantom and enter it in the text box located in the DISTORTION submenu: 3. Clip distortion phantom A onto the X-ray detector cover. 4. Make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in Sidexis. Service routine S010.12 is displayed on the control panel. 5. Press the R key to move the unit back to the starting position. 6. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal for end of exposure (double beep) sounds. 7. The acquired images are transferred from the unit to the Reconstruction Control Unit (RCU). This may take 2 − 3 minutes. The evaluation of the iris calibration is then displayed in the message window. 61 16 813 D3437 D3437.103.01.04.02 i NOTE If the information in the message window indicates that calibration was not successful, check the calibration phantom to make sure that it is not damaged. If the phantom checks out OK (all balls are present and correctly positioned), repeat the procedure starting with point 3 as often as required until the calibration is OK. If you have repeated the procedure three times and still have not attained a positive result, please contact the SIRONA Customer Service Center. 8. If calibration was successful, click SAVE VALUES to save the calibration. • • • Remove the distortion phantom. Distortion calibration is now completed. Continue the calibration procedure with the geometric calibration. 31 9 Adjusting / calibrating the system 9.9 Geometric calibration Sirona Dental Systems GmbH Maintenance Instructions GALILEOS 3. A 1. 2. 4. 5. R WAIT! UNTIL UNIT IS IN STARTING POSITION 6. 9.9 Geometric calibration 1. Go to the GEOMETRY sub-menu. 2. Read off the serial number of geometric phantom A from the ID label of the phantom and enter it in the text box located in the GEOMETRY submenu. 3. Insert geometric phantom A in the bite block holder of the unit. i NOTE Make sure that the phantom is firmly seated and straight in the bite block holder. 4. Make SIDEXIS XG ready for exposure: Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in Sidexis. Service routine S010.13 is displayed on the control panel. 5. Press the R key to move the unit back to the starting position. 6. Press the release button. Hold down the release button until image acquisition is completed and the acoustic signal for end of exposure (double beep) sounds. 32 7. i The acquired images are transferred from the unit to the Reconstruction and Control Unit (RCU). This may take 2 − 3 minutes. The evaluation of the calibration is then displayed in the message window. NOTE If the information in the message window indicates that calibration was not successful, check the calibration phantom to make sure that it is not damaged. If the phantom checks out OK (all balls present are and correctly positioned), repeat the procedure starting with point 3 as often as required until the calibration is OK. If you have repeated the procedure three times and still have not attained a positive result, please check the mechanical geometry of the unit (see Service Manual). Adjust the unit if necessary and then repeat the calibration. If this still does not lead to a positive result, please contact the SIRONA Customer Service Center. 8. • • If calibration was successful, click SAVE VALUES to save the calibration. Remove the geometric phantom. Calibration of the GALILEOS is now completed. 61 16 813 D3437 D3437.103.01.04.02 10 Checking the laser 9.9 Geometric calibration Sirona Dental Systems GmbH Maintenance Instructions GALILEOS d. 10 Checking the laser 1. 2. båÖäáëÜ A B 1. Place the bite block A in the bite block holder for the unit (see User Manual). 2. Check the laser: • Switch on the light localizer with key B on the Easypad. The Lasers generate a red line. Label WARNING Class 2 radiation is emitted during installation. Always keep eyes a minimum distance of 100mm away from the laser. Do not stare into the beam. SAG laser (vertical line) The laser beam must be visible in the center of the bite block as a vertical line. i NOTE To adjust the lasers, see Service Manual No controls are available to adjust the laser power. 61 16 813 D3437 D3437.103.01.04.02 33