 7. CONTROLS AND SETTINGS GB
7.3.2. Travelinmaximumheight:
Respect the warnings. Spring system can cause injury and damage. The settings must be adjusted only on systems with surgical lights installed. 
Remove the screws  and the cover . 
Ease the spring arm by applying a force down with a hand until you see the adjustment nut  in the opening  on the side. 
To increase the height, place a metal rod in the adjustment openings  and turn the nut down. 
To decrease the height, place a metal rod in the adjustment openings  and turn the nut  up. 
Check the arm height by reducing gradually the force exerted on it. Edition 2012 March Page 42 of 59