Servicing Firmware Updates DO NOT point the photomultiplier tube at a light source. This can permanently damage the tube. ▲ DO NOT attempt to clean the mirrors in the optical bench. These mirrors do not come in contact with the sample gas and should not be cleaned. Cleaning the mirrors can damage the mirrors. ▲ Handle all printed circuit boards by the edges. ▲ Do not remove the LCD panel or frame from the LCD module. ▲ The polarizing plate is very fragile, handle it carefully. ▲ Do not wipe the polarizing plate with a dry cloth, it may easily scratch the plate. ▲ Do not use alcohol, acetone, MEK or other Ketone based or aromatic solvents to clean the LCD module, use a soft cloth moistened with a naphtha cleaning solvent. ▲ Do not place the LCD module near organic solvents or corrosive gases. ▲ Do not shake or jolt the LCD module. ▲ Figure 7–1. Properly Grounded Antistatic Wrist Strap Firmware Updates The firmware can be updated by the user in the field via the serial port or over the Ethernet. This includes both the main processor firmware and the firmware in all low-level processors. Refer to the iPort manual for the firmware update procedure. Accessing the Service Mode If the Service Menu is not displayed on the Main Menu, use the following procedure to display it. to scroll to Instrument Controls > 1. At the Main Menu, press press > to scroll to Service Mode > and press . The Service Mode screen appears. 7-2 Model 43i High Level Instruction Manual Thermo Fisher Scientific