Download Model 146i Multi-gas Calibrator - User Manual
Transcript
Calibration Photometer Calibration Photometer Calibration The ozone generator used on the Model 146i qualifies as a transfer standard under current U.S. Environmental Protection Agency requirements. In order to use it as a transfer standard, the user must certify the generator, following the procedures published in the technical assistance document for the calibration of ozone transfer standards. In order to determine a calibration curve for the ozone generator, an ozone primary standard must be used. If an ozone analyzer is also available, the generation of the calibration curve becomes easier. A procedure for both is given below. Calibration Using a Model 49i PS Use the following procedure to calibrate the Model 146i using a Model 49i PS. Refer to the Photometer Calibration Flow Schematic (Figure 4-1). 1. Connect the 49i PS OUTPUT to the 146i EXTERNAL SAMPLE port. 2. Set flow of the 49i PS to the desired level. 3. Determine the ozone concentration at different ozone level settings on the 146i. 4. Determine a calibration curve by plotting the results of Step 3. 5. Change to the 146i Photometer Calibration Span. 6. For highest confidence, reconnect the 49i PS to the 146i and recheck the ozone calibrations. 4-2 Model 146i Instruction Manual Thermo Fisher Scientific