Transcript
2 Site Requirements and Specifications Site Requirements Site Requirements A suitable environment is important to ensure optimum performance of the instrument. Power Consideration The autosampler power supply has wide-ranging capability (see Table 4 on page 27). Consequently there is no voltage selector in the rear of the autosampler. There are also no externally accessible fuses, because automatic electronic fuses are implemented in the power supply. The thermostatted autosampler comprises two modules, the standard or preparative autosampler and the thermostat (G1330B). Both modules have a separate power supply and a power plug for the line connections. The two modules are connected by a control cable and both are turned on by the autosampler module. WA R N I N G Damaged electronics Disconnecting or reconnecting the sampler to thermostat cable when the power cords are connected to either of the two modules will damage the electronics of the modules. ➔ Make sure the power cords are unplugged before disconnecting or reconnecting the sampler to thermostat cable. WA R N I N G Incorrect line voltage at the instrument Shock hazard or damage of your instrumentation can result, if the devices are connected to a line voltage higher than specified. ➔ Connect your instrument to the specified line voltage. 24 Agilent 1200 Series Autosampler User Manual