Download Signet 8058 i-Go™ Signal Converter

Transcript
English
Signet 8058 i-Go™ Signal Converter
3-8058.090
S3
DA L
TA
t 80
English
1
2
3-8058.610D
3
Loop1 +
Loop1 -
N/C
6
Loop2+ 4
Loop2 - 5
GND
9
S3 L
7
S3 L
Output
8
N/C
X
+5VDC
4-20 mA
Input
3
C MA
58
9
Red
Black
)
ck
ield
Sh
Bla
to
(Tie
+GF+
4-20 mA input
Signet 8058 i-Go™
4-20 mA to S3L Converter
S3L Output
Output
S3L
Rev. B 9/05
Input
4-to-20 mA
3-8058.090
ne
N /C
7
8
4-20mA to S
Loo
p2
PW
R
Sig
+GF+
Loo
p1
PW
R
8058 i-Go™
L Converter
35 VD
+G
F+
3
2
1
6
5
4
)
ite
ield Wh
Sh
to
ie
(T
)
(Ground
White
Red
Bla
ck
Description
The 8058 Signal Converter accepts any 4 to 20 milliamp signal and converts it into the serial data format used by the Signet 8250,
8350, 8450 and 8900 instruments known as S3L. (Signet Serial Sensor Link)
•
When used with the 8900 Multi-Parameter Controller, all programming uses the familiar 8900 setup menus.
•
When used with ProcessPro transmitters, the 8058 is configured at the factory to the user's specifications.
•
The 8058-1 wire-mount version is easily mounted anywhere in the interconnnecting wiring between a single 4-to-20 mA source and
the instrument.
•
The 8058-2 DIN-rail mounted version converts two separate 4-to-20 mA source signals into the S3L format.
General
Input:
3-8058-1 Signal Converter, wire-mount
16 mm
(0.63 in.)
Input
4-to-20 mA
Input range:
Output:
3-8058-2:
Accuracy:
Resolution:
Update rate:
Temp. Drift:
4 to 20mA current loop(s), passive
(external power required)
3.6 to 22.1 mA
Digital (S3L) output
calibrated mA
±32 µA @ 25°C
<16 µA
500 mS
±1 µA per °C, max.
Dimensions
+GF+
Signet 8058 i-Go™
4-20 mA to S3L Converter
S3L Output
4-20 mA input
Electrical
Power req.:
4.5 to 6.5 VDC @ < 3.0 mA
Max. voltage:
35 VDC
Max. current:
40 mA
Isolation:
Up to 48 Volts AC/DC
Voltage drop:
5 VDC max.
Reverse polarity protected
104mm
(4.1 in.)
99mm (3.90 in.)
Environmental
Operating Ambient Temperature:
-10°C to 55°C (14°F to 131°F)
Storage Temperature:
-20°C to 85°C (-4°F to 185°F)
Relative Humidity:
•
3-8058-1: 0 to 100%, condensing
•
3-8058-2: 0 to 90%, non-condensing
Shipping Weight:
•
3-8058-1:
0.12 kg (0.26 lb.)
•
3-8058-2:
0.15 kg (0.35 lb.)
Standards and Approvals
•
CE
•
Emissions and Immunity per EN 61326:
EN61000-6-2 Immunity
EN55011 Class B
Emissions
114.5mm (4.51 in.)
Cable:
3-8058-1:
400 mm (15 in.) input,
200 mm (8 in.) output
3-8058-2:
No cable provided
(customer supplied)
Max. recommended cable extensions:
•
Loop in:
300m (1000 ft.)
per S3L guidelines
•
S3L out:
Output
S3L
Specifications
3-8058-2 Signal Converter, DIN-rail mount
22.5mm
(0.89 in.)
Wiring: 8058-1 Wire Mount to Signet 8900
Wiring Tips:
•
Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with the signal.
•
Routing cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
•
Seal cable splices tightly to prevent moisture damage.
•
The cable length from the 8058 to the 8900 must not exceed the S3L limitations as defined in the 8900 instruction
manual.
1
2
3
4
5
6
+GF+
SHIELD
BLACK (5 VDC)
Signet 8058 i-Go™
4-20 mA to S3L Converter
4-20 mA input
S3L Output
RED: (Loop +)
4 to 20 mA
Loop Source
Output
S3L
CAUTION
Loop power must not exceed 35 VDC.
BLACK (Loop -)
Input
4-20 mA
Loop Power
VDC
7
Tie SHIELD to BLACK
+
8
RED (S3L)
9
10
WHITE (GROUND)
11
Tie SHIELD to WHITE
+
-
12
13
14
8900 terminals
1.
2.
3.
Wire the 8058 between the 4-20 mA loop source and the 8900 S3L input terminals as illustrated.
In the 8900 menu, assign a channel for this input, and specify the input parameter.
If the 4-20 mA SOURCE is one of the standard 8900 input parameters (flow, pH, Conductivity, etc.) select that channel type.
If the 4-20 mA SOURCE is not included in the 8900 menu, select OTHER (4-20) and then set additional labels and abbreviations
per the 8900 instructions.
Follow complete instructions in the 8900 manual to set up the new channel.
Wiring: 8058-1 Wire Mount to Signet 8250, 8350, and 8450 ProcessPro Transmitters
When used in conjunction with the Signet ProcessPro Pressure, Temperature or Level Transmitters, the 8058 serves to expand the
operating range beyond the specifications of sensors available directly from Signet.
1.
2.
3.
Order the 8058 according to the instructions on page 4. The 8058 MUST be factory configured for these applications.
Connect the 8058 between the 4-20 mA loop source and the ProcessPro S3L input terminals as illustrated.
Follow the complete CALIBRATION instructions in the ProcessPro manual.
e)
hit
dt
iel
h
eS
Shield
+GF+
4-20 mA input
ld
Signet 8058 i-Go™
4-20 mA to S3L Converter
S3L Output
Output
S3L
RED (S3L)
Input
4-to-20 mA
BLACK (5 VDC)
i
(T
ie
Sh
oW
WHITE (GROUND)
RED (S3L)
BLACK (5 VDC)
(Tie Shield to Black)
CAUTION
Loop power must not exceed 35 VDC.
d
iel
Sh
4-20 mA input
Signet 8058 i-Go™
4-20 mA to S3L Converter
S3L Output
Output
S3L
SHIELD
Input
4-to-20 mA
BLACK (LOOP -)
+GF+
(Tie Shield to Black)
2
d
iel
Sh
+GF+
4-20 mA input
Signet 8058 i-Go™
4-20 mA to S3L Converter
S3L Output
Output
S3L
Input
4-to-20 mA
SHIELD
RED (S3L)
Tie Shield to White
RED (LOOP +)
BLACK (LOOP -)
WHITE (GROUND)
BLACK (5 VDC)
(Tie Shield to Black)
7
6
5
8350-3
8450-3
Terminals
Tie Shield to White
RED (LOOP +)
8350-1
8450-1
Terminals
WHITE (GROUND)
RED (S3L)
BLACK (5 VDC)
8250-2
8350-2
8450-2
Terminals
13
12
11
8250-3
Terminals
16
18
15
17
14
16
13
15
12
14
11
13
8058 i-Go™ Signal Converter
Wiring: 8058-2 DIN Rail Mount to Signet 8900 Multi-Parameter Controller
The 8058-2 accepts two 4 to 20 mA loop input signals and converts them into two S3L serial data signals for input to the Signet 8900,
where each loop is completely independent. The 8058 transmits information to identify the two signals by using the serial number of
the 8058 converter:
•
The manufacture date of the 8058 is included in the serial number
label on the instrument.
The 8058 converter transmits the serial number onto the S3L bus as
the id for the Loop 1.
The 8058 transmits a second serial number for Loop 2 by by adding
12 to the month. (6040630.... becomes 6041830....)
•
•
1.
2.
Loop 1: X
6 04 06 30 1234
Disregard
Year Month Day
Sequential
mfg mfg mfg
Loop 2: X
6 04 18 30 1234
Wire the 8058 between the 4-20 mA loop source and the 8900 S3L input terminals as illustrated.
In the 8900 menu, assign a channel for this input, and specify the input parameter.
If the 4-20 mA SOURCE is from one of the standard 8900 input parameters (flow, pH, Conductivity, etc.) select that channel type.
If the 4-20 mA SOURCE is a type not included in the 8900 menu, select "OTHER (4-20)" and then set additional labels,
abbreviations and units per the 8900 instructions.
1
+GF+
3 2 1
6 5 4
2
+GF+ Signet 8058 i-Go™
3
Two 4-20 mA loop signals can
be connected to the 8058-2.
4-20mA to S3L Converter
S3L
Output
4
5
N/C
6
Loop2
PWR
Loop2 - 5
7
Loop2+ 4
·The red Loop LEDs indicate
when one or both of the 4 to 20
mA loop inputs are powered.
8
BLACK (5 VDC)
RED (S3L)
9
WHITE (GROUND)
10
- 4-20 mA Loop 2
Source
+
6
35VDC MAX
Loop1
PWR
4-20 mA
Input
7
+5VDC
N/C
3
8
S3 L
Loop1 -
2
9
GND
Loop1 +
1
- 4-20 mA Loop 1
Source
+
11
12
3-8058.610D
S3L
DATA
13
·The green DATA LED
indicates when the data
output is active.
7 8 9
+
Power
Supply
12-24 VDC
8900 terminals
Signet 8058
N/C
-
14
The S3L output terminals carry
the input loop signals to the
8900.
CAUTION
Loop power must not exceed 35 VDC.
Wiring: 8058-2 DIN Rail Mount to Signet 8250, 8350, and 8450 ProcessPro Transmitters
When using the 8058-2 with the Signet 8250-3, 8350-3 or 8450-3, both loop sources must be of the same TYPE, and the 8058-2 MUST
be factory-configured to the correct range.
1.
2.
3.
CAUTION
Loop power must not exceed 35 VDC.
Order the 8058-2 according to the instructions on page 4.
The 8058 MUST be factory configured for these applications.
Connect the 8058 between the 4-20 mA loop source and the
ProcessPro S3L input terminals as illustrated.
Follow the complete CALIBRATION instructions in the
ProcessPro manual.
16
15
4-20mA to S3L Converter
S3L
Output
4-20 mA
Input
N/C
Loop2 - 5
8250-3
Terminals
18
17
14
16
13
15
12
14
11
13
6
Loop2+ 4
BLACK (5 VDC)
RED (S3L)
WHITE (GROUND)
7
+5VDC
N/C
3
8
S3 L
Loop1 -
2
9
GND
Loop1 +
1
3-8058.610D
35VDC MAX
8350-3
8450-3
Terminals
+GF+ Signet 8058 i-Go™
-
+
4 - 20 mA
Loop 2 Source
+
4 - 20 mA
Loop 1 Source
+
Power Supply
12 - 24 VDC
8058 i-Go™ Signal Converter
3
IMPORTANT! SPECIAL INFORMATION for ProcessPro Users:
The following information is REQUIRED with all converters that will be connected to:
Signet 8250 Level Transmitter
Signet 8350 Temperature Transmitter
Signet 8250 Level Transmitter
1.
Select the 8058 model preferred:
______ 3-8058-1S Wire-mount current to S3L converter (for use with 8250, 8350 or 8450 -1 or -2 model Transmitters).
______ 3-8058-2S DIN rail mount current to S3L converter, dual input (for use with 8250-3, 8350-3 or 8450-3 Transmitters)
2.
Specify the Signet ProcessPro model to be connected to the 8058:
______ 8250 Level Transmitter
3.
______ 8350 Temperature Transmitter
______ 8450 Pressure Transmitter
Specify the operating range and engineering units from the input 4-20 mA loop source:
Loop 1:
4 mA = _______________(specify units)
20 mA = _______________(specify units)
Loop 2 (-3 models only):
4 mA = _______________(specify units)
20 mA = _______________(specify units)
Note: For dual-loop systems, both input loops must be the same parameter, but the ranges may be different.
4.
Contact a Signet representative
Due to the customization requirements, the 8058 Converter requires special handling when intended for use with ProcessPro
transmitters. Consult with Signet product specialist for these requirements.
Ordering Information
Mfr. Part No.
3-8058-1
3-8058-2
Code
159 000 966
159 000 967
Description
Wire-mount current to S3L converter for use with Signet 8900
DIN rail mount current to S3L converter, dual input for use with Signet 8900
NOTE: If the 8058 converter is to be used with any 8250, 8350 or 8450 ProcessPro Transmitter, add S to the order number, complete
the information noted above, and contact your Signet representative.
S3L Output
Bla
ck
3-8058-1 Wire-mount Current-to-S3L Converter
Lo
op
PW 2
R
Sig
net
S3
DATL
A
3-8058.610D
3
Loop1 -
N/C
7
8
9
S3 L
Loop1 +
2
1
X
GND
MA
+5VDC
6
DC
Loop2+ 4
Loop2 - 5
N/C
4-20 mA
Input
3
Lo
op
PW 1
R
S3 L
Output
)
ite
ield o Wh
Sh
t
ie
(T
d)
(Groun
White
Red
4-20mA to S
4-20 mA input
Signet 8058 i-Go™
4-20 mA to S3L Converter
+GF+
)
ck
ield
Sh
Bla
to
( Tie
Input
4-to-20 mA
Black
+GF+
Output
S3L
Red
8058 i-Go™
L Converter
35 V
+G
F
3 +
2
1
6
5
4
80
N /C
7
8
58
9
3-8058-2 DIN-mount Current-to-S3L Converter
George Fischer Signet Inc., 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090
3-8058.090 Rev. B 8/05 English
© George Fischer Signet 2004
Printed in U.S.A. on recycled paper