Figure 7.4 - Correlation between the breakdown strength and the gap distance [4].
In impulse tests, the sphere-gap is usually protected by a resistance R < 300 Ω.
Correction factors.
As indicated in IEC 60052, when the atmospheric conditions are not the standard as
defined in this section, small corrections are made for air pressure, temperature and
humidity.
Thus, there are two correction factors: for air density (temperature and pressure) and air
humidity.
The correction factor for air humidity is usually small, below 2%, and, as recommended by
Kreuger [4], measurements shall not be made over 90% of relative humidity. The results
obtained may be unreliable due to condensation of water at the sphere surfaces.
- Air density correction factor.
Disruptive discharge voltages corresponding to a given sphere-gap spacing S under
atmospheric conditions other than those specified as standard are obtained by multiplying
the values in tables 2 and 3 of the Standard by a correction factor corresponding to the
relative air density δ.
The relative air density δ is defined by:
δ=
b 273 + t0
×
b0 273 + t
(7.1)