Transmission of results over Modbus
10.2 Changing MODBUS parameters
10.2.2.3
Result values
Result parameters
Values / value range
Description
Sample1.AnalysisID
Number representing the
change for the last result of this
sample stream.
Sample1.InjectionTime
Date and time at which the
injection took place.
Sample1.OEM1.CalOrAnalysisF 
lag@3059

1 for all valve streams
0 for the calibration stream
Sample1.OEM1.NewDataFlag@ value != 0
3058
Value to differentiate between
measurement and calibration.

New data present.

Correct if a new result is
available for this stream.

Writable: can be deleted to 0
by the reader "Modbus
master".
Sample1.Status.Alarm
value != 0
Correct if an alarm occurs
during the analysis.
Sample1.Status.DataNotValid
value != 0
Correct if the result is invalid.
Sample1.Status.Revision
value != 0
Correct if the revision bit switch
was set for this result. The
revision analysis does not
contribute to the mean values.
Sample1.Status.Warning
value != 0
Correct if a warning has
occurred during the analysis.
This value is not currently set by
the device.
Sample1.100PercentNorm
Sample1.Status.OutOfRange
Standardization factor, see
Section "Calculation formulae
(Page 102)".
value!=0
Is correct if the result is outside
the expected range.
CVControl operating software
118
Operating Manual, 11/2012, A5E01428676-03