Download SITRANS CV
Transcript
Transmission of results over Modbus 10.2 Changing MODBUS parameters 10.2.2.3 Result values Result parameters Values / value range Description Sample1.AnalysisID Number representing the change for the last result of this sample stream. Sample1.InjectionTime Date and time at which the injection took place. Sample1.OEM1.CalOrAnalysisF lag@3059 1 for all valve streams 0 for the calibration stream Sample1.OEM1.NewDataFlag@ value != 0 3058 Value to differentiate between measurement and calibration. New data present. Correct if a new result is available for this stream. Writable: can be deleted to 0 by the reader "Modbus master". Sample1.Status.Alarm value != 0 Correct if an alarm occurs during the analysis. Sample1.Status.DataNotValid value != 0 Correct if the result is invalid. Sample1.Status.Revision value != 0 Correct if the revision bit switch was set for this result. The revision analysis does not contribute to the mean values. Sample1.Status.Warning value != 0 Correct if a warning has occurred during the analysis. This value is not currently set by the device. Sample1.100PercentNorm Sample1.Status.OutOfRange Standardization factor, see Section "Calculation formulae (Page 102)". value!=0 Is correct if the result is outside the expected range. CVControl operating software 118 Operating Manual, 11/2012, A5E01428676-03