Download ELx800™ - Probiotek
Transcript
114 | Chapter 4: Performance Verification and IQ/OQ/PQ Procedures To enter the calibration values into KCjunior™: 1. First define the Absorbance Test Plate to be used for the test. To do this, open the Utilities menu, and select Diagnostics|Universal Plate Test. 2. Click New Data Sheet. Click the Help button for guidance. To enter the calibration values into KC4™: 1. First define the Absorbance Test Plate to be used for the test. To do this, open the System menu and select Diagnostics|Define Universal Test Plate| Add Plate. Click the Help button for guidance. Running the Absorbance Plate Test IMPORTANT: Before running the Absorbance Plate Test, ensure that the reader is not running in Rapid Mode. To check this, select UTIL READ and then cycle through the prompts until READ IN RAPID MODE? is displayed. Choose NO for an accurate result. To run the Absorbance Plate Test from the instrument keypad: 1 Start at the Main Menu and select UTIL TESTS CALPLATE. The SELECT TEST menu will appear: S E L E C T T E S T : S Y S T E M C H K S U M 2 Select the appropriate wavelength at the CALIBRATION FILTER screen, and press ENTER to save the value and continue. C A L I B R A T I O N 4 0 5 3 C A L P L A T E 4 5 0 F I L T E R : 4 9 0 4 0 5 * M O R E When prompted, insert the Test Plate into the plate carrier, and press the READ key to begin the test. The Calibration Plate Analysis Report (Figure 19) will be sent to a printer when the test is run. Bio-Tek Instruments, Inc.