Retractor Kit LLPR02 A Instructions for cleaning, sterilization and maintenance 2.2.3. Automatic Cleaning Completely submerge the device in an enzyme solution (Aniosyme DD1, 0.5%) and allow to soak for 5 minutes at room temperature (20-25°C) with ultrasonic vibrations. Use a soft bristled, nylon brush to gently scrub the device for 2 minutes at room temperature (20-25°C) and until all visible soil has been removed. Do not use metallic brush. Particular attention must be given to crevices, lumens, mated surfaces, connectors and other hard-to-clean areas. Thoroughly flush lumens, holes and other hard-to-clean areas with the enzyme solution (Aniosyme DD1, 0.5%). Rinse device in purified water for at least 1 minute. Thoroughly flush lumens, holes and other hard-to-clean areas. Place the devices in a suitable washer/disinfector basket and process through a standard washer/disinfector cleaning cycle. Visually inspect the device for remaining soil and repeat the steps above if necessary. Example of Vario TD cleaning cycle: HAMO ECOLINE LM-25 Cycle Vario TD liquide NEODISHER FLA Pre-cleaning <45°C 2 minutes Cleaning 55°C 5 minutes Neutralization 2 minutes Rinsing Tap water 2 minutes cold water Thermal 90°C 5 minutes Disinfection Demineralized water Drying 22 minutes DO NOT CLEAN THE SINGLE-USE PRODUCTS. 23