Download NI X Series Multifunction Data Acquisition - Data Sheet
Transcript
±100 pA Input bias current Crosstalk (at 100 kHz) –75 dB Adjacent channels –90 dB Non-adjacent channels 1.2 MHz Small signal bandwidth (–3 dB) 4,095 samples Input FIFO size 4,095 entries Scan list memory Data transfers DMA (scatter-gather), programmed I/O NI PCIe/PXIe-6341/6343 USB Signal Stream, programmed I/O NI USB-6341/6343 Overvoltage protection (AI <0..31>, AI SENSE, AI SENSE 2) ±25 V for up to two AI pins Device on ±15 V for up to two AI pins Device off ±20 mA max/AI pin Input current during overvoltage condition Settling Time for Multichannel Measurements Accuracy, full scale step, all ranges 2 μs convert interval ±90 ppm of step (±6 LSB) 3 μs convert interval ±30 ppm of step (±2 LSB) 5 μs convert interval ±15 ppm of step (±1 LSB) None Analog triggers Typical Performance Graph Analog Output Number of channels 2 NI 6341 4 NI 6343 16 bits DAC resolution ±1 LSB DNL 16 bit guaranteed Monotonicity Maximum update rate 900 kS/s 1 channel 840 kS/s per channel 2 channels 775 kS/s per channel 3 channels 719 kS/s per channel 4 channels 6/18 www.ni.com