Maintenance 7. Reassemble pipette top. Attach pipette top to the main housing. Ensure flex cable is behind the top of the pipette trigger and the trigger contacts the limit switch properly (see Figure F.3.1Ae). Fit the edges of the pipette top snugly within the edges along the top of the pipette housing. Ensure that the cables to the power jack are inside the main housing well below the edges to avoid pinching the cables. Hold assembly together firmly while resecuring the two screws. Figure F.3.1Ae: Assembling Pipette Top 8. Charge battery. Charge the new battery before using the pipette (see Section F.4, Recharging Batteries). F.3.1B Installing/Replacing Multichannel Electronic Pipette Batteries The Matrix Electronic pipette is shipped with Nickel Metal Hydride batteries packaged separately. Install the batteries and recharge them for 14-16 hours. Use the following procedure to install the batteries, then continue to Section F.4, Recharging Batteries. Figure F.3.1Ba: Removing Battery Case 46